60 of 92 — 26-0008-1100/5
(Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)
will bring up a warning screen. This warning screen provides
instructions should communication with the Pulse Generator become
impossible. It also serves as notification that the next time the Pulse
Generator is programmed, treatment time per day will automatically
be set to 24 hours per day.
Treatment time per day and treatment start delay time are
nonprogrammable
.
Version 8.1 VNS Programming Software does not support the Model
100B Pulse Generator and the Model 200 Programming Wand. If a
handheld computer is upgraded to Version 8.1 VNS Programming
Software from a previous version of VNS Programming Software,
previous records of Model 100B will not be completely transferred.
The parameter setting information will be available, but the model
number will be blank, the serial number will be shown as ****, and
the End of Service status will be shown as "No" instead of "N/A".
Only use a serial cable provided by LivaNova.
FlashCards are generally not interchangeable! Do not insert the
FlashCard from one handheld computer into any other handheld
computer, unless directed to do so by LivaNova Clinical Technical
Support.
FlashCards should only be inserted in the handheld computer with the
label side up. Figure 36 shows the correct orientation for FlashCard
insertion.
Figure 36.
Correct FlashCard Orientation