Physician’s Manual — VNS Therapy Programming Software (8.1)
6. V
IEW
L
AST
P
ARAMETERS
_____________
You can view the parameter settings from the most recent Pulse Generator
interrogation via the VIEW LAST PARAMETERS Screen (see Figure 33
on page 53). This screen can be accessed from the MAIN MENU screen.
Figure 33.
VIEW LAST PARAMETERS Screen
Tapping the Patient, Device, or Magnet buttons causes the respective
patient, device, or magnet data to be displayed. Using the scroll bar on the
right-hand side of the device can also access the data.
Caution:
Magnet
activation time-stamps
and total activation
counts may be
duplicated for Model
105 Pulse Generator.
See “Precautions ” on
page 59 for more
details.
For Model 105 Pulse Generator, using the cross pattern swiping technique
for magnet activations may cause duplication in the magnet activation
time-stamps and in the total magnet activation counts. Depending on
Generator and Magnet orientation, one or two entries may be shown in the
Magnet History data for each cross pattern swipe. This is an expected
occurrence due to the Pulse Generator design and is not considered a
device malfunction. For more information, see the Pulse Generator
Physician’s and Patient’s manuals.