22 of 92 — 26-0008-1100/5
(Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)
Note:
LivaNova
recommends that during
the initial programming,
the output current be
programmed at 0 mA
and then slowly
increased by 0.25 mA
increments until the
patient feels the
stimulation at a
comfortable level. Even
patients who are
receiving replacement
generators should be
started at 0 mA Output
Current, followed by
incremental increases of
0.25 mA to allow for
reaccommodation to the
therapy.
Figure 10.
PARAMETER SETTINGS POP-UP Menu
2. Select the new target value for the Output Current by tapping the
desired value on the screen. If the target value selected is greater than
0.25 mA above the currently programmed value setting in the Pulse
Generator, an Output Warning will appear (see Figure 11 on page 22).
Figure 11.
OUTPUT WARNING