Making Temperature Measurements 

Place temperature probes in positions to take desired measurements. When measuring surface temperatures, solid mechanical 
contact is important. 

Use Spindle Error Analyzer (SEA) Software to extract temperature readings. 

 

Block Diagram - Temperature 

 

Connections to Temperature Probes 

The two wires of each temperature probe are connected according to the chart below. The probes are by nature unpolarized so 
there is no differentiation between the two conductors. 

  

Pin 

Numbers Connection 

Pin 

Numbers

Connection 

1, 9 

Probe 1 (T1) 

5, 13 

Probe 5 (T5) 

2, 10 

Probe 2 (T2) 

6, 14 

Probe 6 (T6) 

3, 11 

Probe 3 (T3) 

7, 15 

Probe 7 (T7) 

4, 12 

Probe 4 (T4) 

No connection 

 

TMP190 Temperature Specifications

1

 

 

Accuracy 

±1.8 °F@40 °F – 180 °F 

±1. 0°C@4°C – 8 2°C 

Output Voltage 

±10VDC 

Measuring Range 

40 °F – 180 °F 

4 °C – 82 °C 

Temperature Probe 
Interchangeability Error 

±0.2 °F @ 68 °F 

±0.1 °C @ 20 °C 

1

Shifts as high as 4 °C may occur in high EMI conditions (10 V/m). 

 

1

9

8

15

Summary of Contents for CPL190

Page 1: ...Sensor Systems CPL190 CPL290 CPL490 MM190 TMP190 EN190 Lion Precision 563 Shoreview Park Road St Paul Minnesota 55126 7014 Telephone 651 484 6544 Fax 651 484 6824 www lionprecision com Copyright 2015...

Page 2: ...icators and Connectors 8 Multiple Channel Measurements 8 CPL490 Specifications 9 Changing CPL190 290 490 Bandwidth Settings 10 MM190 Meter and Function Module 11 MM190 Specifications 12 TMP190 Tempera...

Page 3: ...ger Normally when the probe tip is touched the probe stops functioning and voltage is reduced to near zero But under certain circumstances the voltage may cause a slight tingle or burning sensation es...

Page 4: ...put is not guaranteed to be accurate even though the voltage may be in range and continue to change Probes are calibrated to specific modules With multiple channel systems be sure the number on the pr...

Page 5: ...appear valid but the Calibrated Range Indicator will be red Out of Range indicator signals available through the system DAQ connector are activated anytime the Calibrated Range Indicator is red Analog...

Page 6: ...db 15 kHz 10 30 Coarse Zero Adjustment Range 10 VDC Fine Zero Adjustment Range 1 VDC Meets ANSI ASME B5 54 Standard Yes Probe Interchangeability Typical sensitivity variation for same probe model 10 O...

Page 7: ...red indicate that the probe is out of its calibrated range and the output is not guaranteed to be accurate even though the voltage may continue to change Probes are calibrated to specific modules With...

Page 8: ...ion DAQ connector on the rear of the enclosure is differential See the enclosure section of this manual for specifics on the system DAQ connector page 19 Ground A banana plug cable can be connected he...

Page 9: ...e dependent Oscillator Frequency Probe Calibration dependent 6 4 MHz typical Output Impedance 0 Output Max Voltage 13 5 VDC Output Max Current 10 mA 1 These specifications are typical for standard com...

Page 10: ...e following table shows the bandwidths available and their associated dip switch settings Bandwidths listed are approximate Actual values depend on the probe and calibration CPL190 290 S2 CPL490 S1 2...

Page 11: ...idual channel to be displayed A or B or the sum or difference of A and B 3 Select a peak function 4 Select metric or inch units Display Units The MM190 measures the analog voltage from the selected se...

Page 12: ...MM190 BNC connector is scaled the same as the BNC output voltage from the selected sensor channels For example if the sensor channels are scaled at 1 V 1 m the MM190 will also scaled at 1 V 1 m When...

Page 13: ...Lion Precision Spindle Error Analyzer SEA and requires the SEA software to access the temperature signals Operation Temperature Sensing Temperature Probes Thermistor type temperature probes are used w...

Page 14: ...ording to the chart below The probes are by nature unpolarized so there is no differentiation between the two conductors Pin Numbers Connection Pin Numbers Connection 1 9 Probe 1 T1 5 13 Probe 5 T5 2...

Page 15: ...ifferential Inputs Encoder and index inputs can be configured as single ended or differential In single ended mode the signal is measured between the input and ground In differential mode the signal i...

Page 16: ...ncoder Index Detailed datasheet on the 75175 Transceiver is available at http focus ti com lit ds symlink sn75175 pdf 499 499 10K 10K 1 4v Threshold Diffferential Input Configuration Single Ended Inpu...

Page 17: ...mber of module slots Input power Tip up handle Mounting flanges EN191 1 15 VDC No Yes EN192 2 15 VDC No Yes EN193 3 15 VDC No Yes EN196 6 100 250 VAC 50 60 Hz Yes No EN198 8 100 250 VAC 50 60 Hz Yes N...

Page 18: ...s direct connection to the enclosure This supply features a high frequency 100kHz switching supply The high switching frequency allows the sensing modules to operate at maximum resolution DC Output Vo...

Page 19: ...und 23 Analog Out Channel 8 25 Analog Out Channel 7 26 Analog Out Channel 6 28 Analog Out Channel 5 30 Analog Out Channel 4 31 Analog Out Channel 3 33 Analog Out Channel 2 34 Analog Out Channel 1 57 A...

Page 20: ...1 3 33 84 6 mm 2 606 66 2 mm 3 806 96 7 mm EN192 4 73 120 1 mm 4 006 101 8 mm 5 206 132 2 mm EN193 6 13 155 7 mm 5 406 137 3 mm 6 606 167 8 mm 4 2 0 6 0 5 1 8 2 6 7 0 0 4 2 8 1 C inch mm Front View To...

Page 21: ...21 Mechanical Specifications EN196 EN198 Model A EN196 10 1 257 mm EN198 14 3 364 mm inch mm Side View Front View A 12 4 315 5 8 147 3...

Page 22: ...amond accepts and holds a high polish minimizing the possibility of scratching the measured surface Minimal Wear Increased accuracy and longer life Contact Force Adjustment There is a contact force ad...

Page 23: ...23 Mechanical Detail...

Reviews: