remain stable to ensure a very good reproducibility over the complete measuring range. 

IV. Operation 

1. Power on/off 

Power on:

   

Short press the power button to turn on the gauge. Display the version number and serial number, then 
the recorded data of last measurement are displayed after the gauge is turned on.     

Power off:   

Long press the button to shut down the gauge, or the gauge will automatically shut down in 3 minutes 
without any operation. 

2. Device setting 

In the off state, long press the button for 3s to enter the setting interface.After entering the setting interface, 
if there is no operation for more than 20 s, the gauge will automatically shut down.Short press the button 
can select the settings; Long press the button for 3 s and less than 5 s to confirm the settings; long press 
the button for 5 s, the gauge exits the setting and shuts down and the setting is invalid.   

Language settings 

The instrument is available in Chinese, English, Russian, Turkish. 

Setting method: In the main setting interface, short press power button to select "Language" option, long 
press the button for 3 seconds to confirm the selection, short press the button again to select the desired 
language, long press the button for 3 seconds to confirm the selection, and return to the main setting 
interface. 

 

 

                       

  Setting interface                          Language selection 

Measurement modes selection 

There are three measurement modes: Ferrous substrate measurement (Fe), Aluminum substrate 
measurement (Al), and Auto identification (Fe/Al). Under normal circumstances, it is enough to use Fe/Al 
auto identification mode. This mode features identification of the ferrous putty and galvanized iron 
substrate. When the measurement substrate is settled, you can select Fe or Al as the fixed 
measurement mode.   

Summary of Contents for LS236

Page 1: ...Shenzhen Linshang Technology Co Ltd 3 12 2022 Page 1 of 8 LS236 Coating Thickness Gauge User Manual V2 0 Please read this manual carefully before using and reserve it for reference ...

Page 2: ...isible under strong light The top display uses OLED OLED ensures normal display in a low temperature environment of 40 C It also features 3 color backlight display When the device identifies the ferrous putty the backlight turns red In case of the galvanized iron substrate the backlight turns green And under other situations the backlight is white Standards for the product DIN EN ISO 2808 Paints a...

Page 3: ...ent angles 5 Using LCD and OLED as display devices simultaneously LCD ensures that the test results are clearly visible under strong light and OLED ensures normal display in a low temperature environment of 40 C 6 The device can identify the ferrous putty as well as the galvanized iron substrate It also features 3 color backlight display 7 Wear proof ruby probe tip for long term use 8 Dual use pro...

Page 4: ...ess than 5 s to confirm the settings long press the button for 5 s the gauge exits the setting and shuts down and the setting is invalid Language settings The instrument is available in Chinese English Russian Turkish Setting method In the main setting interface short press power button to select Language option long press the button for 3 seconds to confirm the selection short press the button ag...

Page 5: ...rm and back to the main setting interface Main setting interface Unit selection interface Exit In the main setting interface short press power button to select Exit long press the button for 3 seconds to confirm and enter the measurement interface 3 Zero adjustment A zero adjustment is required when using the gauge for the first time after inserting new batteries working with different materials o...

Page 6: ... 3 4 The zero adjustment completed when there is the buzzer sound again and the LCD screen displays 0 0 3 5 After the zero adjustment is completed place the standard film on the plate substrate if the measurement value is stable and deviation from standard value within 3μm the gauge can be used normally Note After the zero adjustment completed when repeating measurement on the same spot the readin...

Page 7: ... than 9 sets of data are stored the earliest recorded value is automatically deleted and No 1 is the last test data Recorded data will not loss after powering off V Precaution 1 The device must be zero adjusted respectively with the iron base and aluminum base zero adjustment plates Otherwise there may be abnormal identifications of the ferrous putty and galvanized iron substrate 2 It is strictly ...

Page 8: ...ally please send the whole gauge to our company for maintenance 2 Provide users with spare parts and lifelong maintenance services 3 Provide the users with the gauge calibration service 4 Free technical support for long term Manufacturer Shenzhen Linshang Technology Co Ltd Website www linshangtech com Service hotline 86 755 86263411 Email sales21 linshangtech com ...

Reviews: