Linseis Chip-DSC How-To Download Page 7

How To Raman spectroscopy with Chip-DSC  

 

4.4 evaluation 

o

 

Start the Linseis Evaluation software  

o

 

Load the measurement and evaluate to your investigations needs 

o

 

Use the Raman-extension and choose the folder where your files are stored 

o

 

Use the data explorer function and enable the image function 

 

o

 

Trace over your measurement and confirm (Enter) at the values you’re interested in 

 

o

 

To find the matching files, the timestamp is shown on your picture 

 

o

 

If you adjusted the spectral data, just rerun the Raman extension to recreate the pictures and 
reload the optimized spectrum 

Summary of Contents for Chip-DSC

Page 1: ...1 How To Raman spectroscopy with Chip DSC Linseis Messgeräte GmbH Gerlach Stand 18 11 2021 ...

Page 2: ...1 General information 3 2 Requirements 3 2 1 Chip DSC 3 2 2 Raman adapter 3 2 3 Raman Spectrometer 4 3 Installation 5 3 1 software 5 3 2 hardware 5 4 Measurements 5 4 1 Linseis Settings 5 4 2 WP Enlighten setting 6 4 3 measurement 6 4 4 evaluation 7 ...

Page 3: ...investigations that cannot all be displayed here For more Information read the other available instructions or specific manuals for the Chip DSC 2 Requirements 2 1 Chip DSC o Check if all necessary parts are delivered and intact o Necessary parts Chip DSC main device Sensor USB cable power supply calibration material SOA PC 2 2 Raman adapter o Depending on the Chip DSC device is a special adaptor ...

Page 4: ... The Raman spectrometer you ve choose comes with all necessary parts you need to perform a spectroscopy Differences in specifications could be caused by individual investigations o Necessary parts Raman spectrometer detector fibre optics cables power supply USB cable ...

Page 5: ...ip DSC at the measurement desk for further information check HowTo Setup Chip DSC o Check if the main functions of the device work properly o Fix the detector at the detector handle and place it at the measurement cell o Place and connect the Raman spectrometer nearby the Chip DSC o Enable the laser in WP Enlighten Photonics software and check position of the laser on the sensor via the camera liv...

Page 6: ...0 C o Adjust Light power by response of the Raman spectrum 20 000 counts are recommended o Adjust the batch measurement settings regarding your investigation count and period o Set Laser Mode Manual to avoid ghost peaks on the baseline 4 3 measurement o Make sure the sample is placed correctly on the sensor and the laser is focused on the sample using the camera Toggle light source o Check if all ...

Page 7: ...and choose the folder where your files are stored o Use the data explorer function and enable the image function o Trace over your measurement and confirm Enter at the values you re interested in o To find the matching files the timestamp is shown on your picture o If you adjusted the spectral data just rerun the Raman extension to recreate the pictures and reload the optimized spectrum ...

Reviews: