Gas Test/Calibration Troubleshooting
Problem
Possible Cause
Solution
No response to gas
Gas Cylinder Empty
Check cylinder gauge. Replace cylinder
Hose blocked or kinked
Check hose and ensure no gas flow restriction
Instrument fails gas test
Gas cylinder empty
Check cylinder gauge. Replace cylinder
Gas cylinder out of date
Check cylinder expiry date. Replace cylinder
Hose blocked or kinked
Check hose and ensure no gas flow restriction
Calibration drifted
Calibrate instrument
Gas flow not started in time
Repeat gas test. Start gas immediately
Instrument fails calibration
Gas cylinder empty
Check cylinder gauge. Replace cylinder
Gas cylinder out of date
Check cylinder expiry date. Replace cylinder
Hose blocked or kinked
Check hose and ensure no gas flow restriction
Calibration drifted
Calibrate instrument
Instrument stabilisation time
too short
Reset instrument stabilisation time using G-TECTA
®
PC
Software
Instrument passes gas test but
will not enter calibration mode
Zero not performed
from options menu
Initiate zero on instrument menu
Instrument incorrectly configured
Re-configure instrument.
Send to Linde Group service centre
G-TECTA
®
4G Product Specification
Dimensions
71 x 114 x 48 mm
Weight
295g (with alligator clip & 3 sensors)
IP Rating (Dusts & Liquids)
IP65 & IP67
Operating Temperature Rating
-20°C Ta +55°C (-4°F Ta +131°F)
Humidity Rating
20-90% RH – non-condensing, continuous operation
Display
128 x 64 Pixel with backlight
Warm-up time period
Approx 45 seconds
Typical response time
T90: 10 seconds (oxygen)
20 seconds (toxic & flammable)
Alarm sounders*
95dB(A) Optional multiple alarm sounds
Visible alarms
Dual colour Blue/Red flashing LEDs
Vibrating alarm
Internal vibrating
Repeatability
±2% FSD, 6 months
Electromagnetic Compatibility
2004/108/EC
Explosion Protection Certification
IECEx/ATEX/CSA
IECEx
IECEx BAS 09.0117
ATEX
BASEEFA09ATEX0245
CSA
CSA22.2, 152
Approval Codes
Global
IECEx II 2G Ex ia d IIC T4 Gb
CSA
Class I Division 1, Groups A, B, C and D
Safety & Operation Standards
EN50014, EN50020, EN50018, 94/9/EC
EN50270, EN50271
* It is recommended that the sounder output is periodically tested to ensure the instrument continues to operate as specified