2
English
Manual Thermometer Limit 60
Digital thermometer with two inputs for measuring with two probes at same time.
Up to 99 measurements can be stored over a period of time, by setting automatic logging
interval or manually stored. Stored measurements can be transfer to PC.
The display shows Max-Min-Average and difference between two probes. Hold function.
Easy to compensate for errors of a probe. Display backlight. Switchable between °C and °F
Warning of low battery capacity. Probes of type K, J, T and E can be used.
Specification
Range type K
°
C
-200 … +1370
Reading
°
C
0,1
Accuracy
°
C
0,8 ± 0,5 %
Operation temperature
°
C
-10 … +50
Storage temperature
°
C
-40 … +60
Dimension
mm
175x85x30
Weight
g
270
Battery
1 pc 9 V 6F22
Saftey informations
CE. Safety EN55022, EN55024
EN61326
Electrical equipment for measurement, control and laboratory use.
Buttons
On / Off
On / Off Display backlight
HOLD
On / Off Display backlight
°
C
°
F K
Select between Celsius Fahrenheit and Kelvin.
MAX/MIN
Toggle between max, min, and average.
T1 T2 T1-T2
Toggle between T1 and T2 and different between T1 and T2.
SEND
Enter USB mode for connection to PC.
STORE
Start logging mode. Memory for 99 readings.
RECALL
Show logged readings.
SETUP
Start and exit setups
Increase and decrease by setup.
ENTER
Confirm setup.
SET UP
Toggle with SETUP button between the different setup options.
Confirm selected setup with ENTER.
Logging interval
Display shows INTERVAL. Select logging interval.
Press with
Interval from 0:01 to max 59:59 minutes.
For manual interval set 0:00.
Probe type
Display shows TYPE. Select type K, J, T or E with
.
Chance Offset
Display shows OFFSET T1. Press with
until the display
shows correct reading for T1 . Repeat for T2.
Summary of Contents for 60
Page 1: ...60 ...