background image

Assembly Instructions

Follow the assembly instruction for the appropriate Liko lift.

Likorall, MasterLift, Golvo

: Fixed assembly

Always use 

one of the sleeves, included with the scale.

1. Remove the lifting accessory from the lift.
2.  Place one sleeve on the scale´s lower attachment point. 

Mount the lifting accessory under the scale, using bolt

 

and locking nut provided. 

3.  Mount the scale on to the lift, using bolt and locking nut 

 

provided.

Viking, Uno: 

Fixed assembly with adapter

Always use 

two sleeves, included with the scale

1. Remove the lifting accessory from the lift.

2.  Mount the appropriate adapter (see previous page) 

 

on to the lift in accordance with the assembly instruction.

3.  Place one sleeve on the scale´s lower attachment point.  

Mount the lifting accessory under the scale, using bolt and 

 

locking nut provided.

4.  Place one sleeve on the scale´s upper attachment point.  

Mount the scale in its upper attachment to the adapter, using 

 

bolt and locking nut provided. 

 When disassembling the scale, the adapter must be removed from the lift 

before a lifting accessory is reassembled on to the lift.

3

www.liko.com

2

www.liko.com

Before assembly, make sure you have the following:

• Tool kit, 2 screws, 2 locking nuts 

• Certificate of EU approval

 

2 sleeves 

• Declaration of Conformity

• Battery 

• Instruction Guide from manufacturer

• User Guide from Liko 

• Declaration of Calibration

0

Weighing with LikoScale 350

The scale can be reset to zero with or without the sling. If the scale is reset to zero 

without the sling attached to the sling bar, the weight of the patient will include the weight 

of the sling. Therefore the weight of the sling must be subtracted to obtain the correct 

weight of the patient. 

Resetting the scale with the sling is described below:

1.  Connect the sling to the sling bar. 
2. Start the scale by pressing the     button.
3. Reset the scale by pressing the     button. 

Dashes will move across the display for a few seconds while the scale is resetting.

When the scale is ready, the value ”

0.0

” will appear on the display. 

   Note! No more than 35 kg (77 lbs) can be tared to zero.

4. Remove the sling from the sling bar and apply the sling to the patient in accordance 

with the instruction guide for the sling and lift. Connect the sling to the sling bar.

5. Raise the sling bar so the patient rises slightly. Never lift a patient higher than  

necessary.

6.  When the lifting motion has stabilized, press the     button. The patient’s weight can 

now be read on the display. The scale turns off automatically after about 60 seconds. 

For further readings, press the 

  

  button.

7.  Lower the patient carefully and remove the sling from the sling bar.

Operation

The normal range for operation of LikoScale 350 is a temperature b5°C and 

+35°C (+41°F and +95°F) at a relative humidity not exceeding 95%. 
The scale must not be stored at temperatures below -20°C or above +50°C (-4°F and 

+122°F). 
When the scale is stored for more than 6 months, the battery should be removed. 

 

Follow instructions for installing/replaceing the battery.
Before weighing, ensure the scale has been installed in accordance with the Assembly 

Instructions on pages 2-3.

 Before lifting, check that the lifting accessory is hanging vertically and can 

move freely. 

 Multirall, Likorall 200 / R2R: 

Quick-release assembly

 Liko lifts equipped with (Q-Link) quick fastener

 

require the 

 

Adapter Kit LikoScale, Prod. No. 3156232
Always use 

one of the sleeves, included with the scale.  

 

 Mount the Adapter Kit LikoScale in accordance with the 

 

assembly instruction.

     

Installing/replaceing the battery

LikoScale 350 is delivered with a 9 V battery with a battery life of about 3000 readings. 

Installation of battery:

1.  Open the door on the back of the scale.

2.  Install the battery. 

3.  Close the door.

When the battery needs to be replaced, a small battery symbol will 

appear in the upper left corner of the display.

NOTE! 

Old batteries should be left at the nearest recycling station.

Sleeve

Sleeve

Sleeve

Sleeve

Reviews: