LikoScale™ 200 accessory  

• 

 7EN160154 Rev 4

3

www.hill-rom.com  I  www.liko.com

The normal climate range for operation of LikoScale 200 accessory is b10

C (50

F) to +40

C (104

F) at a relative 

humidity not exceeding 95%. 
The scale must not be stored at temperatures below -20°C (-4°F ) or above +50°C (+122°F). 
When the scale is stored for more than 6 months, the battery should be removed. Follow instructions for Installing/

replaceing 

the Battery.
Before weighing, ensure the scale has been installed in accordance with the Assembly Instructions on pages 2-3.

 Before lifting, check that the lifting accessory is hanging vertically and can move freely. 

Setting of unit mode (kg/lbs.)

Press the (kg/Ibs) - button on the scale to display either kg or lbs.

Weighing with LikoScale 200 accessory

The scale can be reset to zero with or without the sling. If the scale is reset to zero without the sling attached to the sling bar, 

the weight of the patient will include the weight of the sling. Therefore the weight of the sling must be subtracted to obtain 

the correct weight of the patient. 
Resetting the scale with the sling is described below:
1. Connect the sling to the sling bar. 
2. Start the scale by pressing the ”OPERATE” button.
3.  Reset the scale by pressing the ”ZERO” button. 

 

When the scale is ready, the value ”

0.0

” will appear on the display. 

4.  Remove the sling from the sling bar and apply the sling to the patient in accordance with the instruction guide for the sling 

and lift. Connect the sling to the sling bar.

5. Raise the sling bar so the patient rises slightly. Never lift a patient higher than necessary.
6.  When the lifting motion has stabilized, press the ”OPERATE” button. The patient’s weight can now be read on the display. 

The scale turns off automatically after about 60 seconds. For further readings, press the ”OPERATE”

 

button.

7. Lower the patient carefully and remove the sling from the sling bar.

Operation

Sleeve

Sleeve

Alt. 1

Alt. 2

Likorall

 overhead lift, MasterLift overhead lift, Golvo

 mobile lift: Fixed assembly

1.  Remove the lifting accessory from the lift.
2.   Attachment point under the scale:

 

Alternative 1

 with Universal sling bar  

 

  Place one sleeve on the scale´s lower attachment point.

 

  Mount the lifting accessory under the scale, using bolt and locking nut provided.

 

Alternative 2

 with Standard sling bar

 

  Mount the lifting accessory under the scale, using bolt and locking nut provided.
3.   Mount the scale on to the lift, using bolt and locking nut provided.

Multirall

 overhead lift, 

Likorall

 overhead lift, MasterLift overhead lift, 

Golvo

 mobile lift: Quick-release assembly

 Liko lifts equipped with (Q-Link) quick fastener

 

require the Adapter Kit LikoScale, Prod. No. 3156232.

1.  Remove the lifting accessory from the lift
2.   Place one sleeve on the scale´s upper attachment point. 
3.   Mount the Adapter Kit LikoScale in accordance with the assembly instruction. 
4.  Mount the scale on to the lift.
5.  Mount the lifting accessory under the scale.