A Liko™ lift system is a combination of products. Each product has its own maximum load marking. The lowest specified 

maximum load of the products that comprises the system sets the maximum load for the complete lift system. 

The lift system can contain following products; lift rails, a lift motor, a sling bar, a sling or other recommended accessories. 

Identify the maximum load for the complete lift system by reading the maximum load marking on each product in the system. 

Contact your Hill-Rom representative for support and advice.

Intended use

A SlingGuard™ sling bar is used as an aid in carrying out lifts but does not come into contact with the patient, therefore we do 

not go into different patient conditions in the instriction for use. Surrounding (functional) environment: Temp. +5 °C to +40 °C, 

relative humidity 15% to 93% non-condensed. Barometric pressure range of 70 kPa to 106 kPa. 

 

In the event of any queries regarding this product, contact your Hill-Rom representative.
For guidance and support in choosing the correct sling, visit www.hillrom.com or contact your Hill-Rom representative.
Note! Always read the instruction for use that comes with Liko´s different sling models for correct and safe use of the sling. 

The total maximum load for a lift system is always determined by the product in the system with the lowest 

specified maximum load.

Read and understand the Instruction for use in its entirety before using the product.

Use of the product without a full understanding of the contents is subject to risk.

Read the instruction for use for both the patient lift and lifting accessories before use. Lifting and 

transferring a person always involves a certain level of risk. It is important to completely understand the 

contents of the instruction for use. The equipment should only be used by trained personnel. Contact your 

Hill-Rom representative in the event of any uncertainties or questions. 

Under no circumstances may the sling bar be modified. In the event of questions, please contact Hill-Rom.

Summary of Contents for 3308020

Page 1: ...450 Prod No 3308020 SlingGuard 450 with Quick Hook Multi Prod No 3308520 SlingGuard 670 Twin Prod No 3308040 SlingGuard 670 Twin with Quick Hook Multi Prod No 3308540 7EN161113 Rev 5 ENGLISH 2020 Sli...

Page 2: ...ed assembly of SlingGuard sling bar to lift 8 Assembly of SlingGuard sling bar with Quick Hook Multi 9 Dimensions 9 SlingGuard 450 sling bars 9 SlingGuard 670 Twin sling bar 9 Cleaning and disinfectio...

Page 3: ...tructions for use before use Date of manufacturing Important Failure to follow the instructions may lead to damage to the product Caution consult instructions for use Read instructions for use before...

Page 4: ...the lift Fixed attachment or the quick attachment device Quick Hook Multi SlingGuard Sling bars has latches to protect against involuntary unhooking of the slings suspension loops Medical Device Clas...

Page 5: ...106 kPa In the event of any queries regarding this product contact your Hill Rom representative For guidance and support in choosing the correct sling visit www hillrom com or contact your Hill Rom r...

Page 6: ...e for the selected sling before lifting Attach lifting accessories correctly so that the sling bar hangs level and all sling bar hooks take all the weight when lifting Before lifting always ensure tha...

Page 7: ...g bar hooks Note For correct and safe use of a sling always study the instruction for use for the selected sling before lifting Sling with 4 suspension loops Upper strap loops Leg support loops A A B...

Page 8: ...t Fits LikoGuard L Overhead Lift Prod No 3301030 LikoGuard XL Overhead Lift Prod No 3301040 Torque 1 25 0 25 Nm 11 2 2 in Ibf 1 2 3 5 4 6 7 Tools required Torque Wrench 0 10 Nm 88 in Ibf Torx T25 X 2...

Page 9: ...b Installation dimensions the distance between the sling bar hooks attachment point for sling etc to the sling bar s central suspension point CSP to lift A B C H 200 757 133 171 A B C H 7 9 29 8 5 2...

Page 10: ...o avoid the risk of infection Necessary precautions must be taken to eliminate all visible stains and dirt The cleaning and disinfection methods described are specifically aimed at the lift and all ac...

Page 11: ...Alcohol Isopropyl alcohol 70 5 0 7 0 Viraguard or equivalent Veridien Quaternary ammonium n Alkyl dimethyl benzyl ammonium chlorides 0 105 n Alkyl dimethyl ethylbenzyl ammonium chlorides 0 105 11 5 12...

Page 12: ...and any required cleaning and disinfection of the medical device after use and prior to its disposal Customers should adhere to all federal state regional and or local laws and regulations as it perta...