background image

5.1 Product Changes, Transfer Motor LR

Prod. No. 3126043, 3126044

2014 

Changes

Revision:

From S/N: 

Project

Dampers are removed from the Transfer Motor. They are only required when the 

motor is installed in a rail system equipped with a Traverse Switch, according to 

the Installation Hand Book.

03

6002818

SUS000602

2012 

Changes

Revision:

From S/N:

Project

Change of material and design on the cover of Transfer Motor to flame class V-0. 

Model is modified for improved resistance against water, for IPX4 demands. 

The 

turning wheel is removed and pressure spring is replaced with new pressure 

spring.

Transfer motor Likorall ES is approved according to 3rd edition of EN60601-1. 

02

6002300

1000

New version of Dampers.

01

6002230

1778

2010 

Changes

Project

Discontinued product; Transfer Motor Likorall 240/242 (Prod No 3126043) 

1392

 

2006 

Changes

From S/N:

Project

New version of Dampers.

6000953

464

Service Manual Transfer Motor

 

• 3EN620401 Rev. 7   

11

 

www.liko.com  

l

  www.hill-rom.com 

Summary of Contents for 3126043

Page 1: ...Traverse 3EN620401 Rev 7 ENGLISH Transfer Motor over head lift accessories Service Manual Prod No 3126043 Transfer Motor Likorall 240 242 overhead lift accessorie discontinued 2013 3126044 Transfer M...

Page 2: ...ailable after last manufacturing date Repairs and maintenance may only be performed by personnel authorized by Hill Rom and using original Liko spare parts For information regarding Service and Mainte...

Page 3: ...l ES overhead lift accessorie 5 1 3 Transfer Motor Traverse 6 2 0 Spare Parts List 2 1 Transfer Motor Likorall overhead lift accessorie 7 2 2 Transfer Motor Traverse 8 3 0 Periodic Inspection Informat...

Page 4: ...e correctly attached to the hooks on the sling bar This should be checked when beginning to lift the patient when the sling straps are taut but the patient is still seated that the patient if using Li...

Page 5: ...IR Prod No 3126060 All models of Likorall ES can be equipped with a wireless hand control IR The hand control with IR control has six buttons and operates as the fixed hand control please see explana...

Page 6: ...gard to the patient s needs Transfer Motor Traverse together with Liko Likorall lift units complies with the requirements of EN ISO 10535 IEC 60601 1 EN 60601 1 2 UL 60601 1 and CAN CSA C22 2 No 601 1...

Page 7: ...r pair 3SE640257 31290124 541 Washer Transfer Motor LR set x12 x4 3SE690242 Spare Parts Positions PROD NO PRODUCT Liko Spare Parts Spare Parts List Overhead Lifts 29 www hill rom com l www liko com 32...

Page 8: ...ir 3SE640249 31090015 326 Electronic card with cover pce 3SE640250 Spare Parts Positions PROD NO PRODUCT Liko Spare Parts Spare Parts List Overhead Lifts 30 www hill rom com l www liko com Transfer mo...

Page 9: ...pproval to use the accessory Approved Not approved If the accessory has one or more inspection points Not approved the product must not be used Action required Actions according to the inspection item...

Page 10: ...d Life Time Normal use is defined as 8 12 normal lifts day Year Transfer Motor over head lift accessories 10 Expected Life Time The expected life time of Liko Transfer Motors are 10 years of normal us...

Page 11: ...the cover of Transfer Motor to flame class V 0 Model is modified for improved resistance against water for IPX4 demands The turning wheel is removed and pressure spring is replaced with new pressure s...

Page 12: ...e Prod No 3126046 2008 Changes S N Project New version of guide wheel 6030073 767 2007 Changes From S N Project New version of electronic card Rev B 6030053 513 Service Manual Transfer Motor 3EN620401...

Page 13: ......

Page 14: ...Liko AB Nedre v gen 100 SE 975 92 Lule Sweden www hill rom com I www liko com Copyright Liko R D AB 2017...