LIGHTMED TruScan Pro
–
Operator’s Manual Rev. No. 01
Page
24
of
109
•
Heat-absorbing
Movement ranges
•
Longitudinal (in/out): 113 mm
•
Lateral (left/right): 108 mm
•
Vertical (up/down): 35 mm
•
Fine movement range: 10 mm
•
Chinrest range: 70 mm
3.2.2
LIO Specifications
Item
Description
Model
Keeler All Pupil II
Retinal spot size
300µm nominal at focus with 20D Laser Lens
Illumination power
From headset with LED illumination. For other LIO models, the
illumination power is also available from laser console. Internal PSU
allows intensity adjustment.
Fiber length (m)
3
Weight
Less than (<) 500 g without laser attachment
Safety filter
Fixed filter OD4
Beam divergence
Cone angle 20°
Item
Description
Model
Keeler Vantage
Retinal spot size
1100µm +/- 20%
Measurement taken at 280mm from the front face of the LIO
Illumination power
From headset with LED illumination. For other LIO models, the
illumination power is also available from laser console. Internal PSU
allows intensity adjustment.
Fiber length (m)
3
Weight
Less than (<) 500 g without laser attachment
Safety filter
Fixed filter OD4
Beam divergence
Cone angle 20°
3.2.3
Endo-ocular Probes Specifications
Item
Description
Probe types
•
Straight
•
Curved
•
Aspirating
Fiber length (m)
3
Safety filters
OD4 in housing needs to be equipped to
operating microscope before the laser can be
fired
Sterile
Sterilized by ethylene oxide and packed as
single-use-device, ready for use
Beam divergence
Less than (<) 0.2 NA