LIGHTMED TruScan Pro
–
Operator’s Manual Rev. No. 01
Page
21
of
109
Section 3
Product Specifications
3.1
Console Laser System Specifications
3.1.1
General Specifications
Item
Description
Electrical input
•
100 to 240 VAC
•
50/60 Hz
•
Single phase
Power
•
150 W
•
12 V @ 15 A
•
24 V @ 7 A
Fuse rating
T8AH250V @ 100 to 230 VAC (time lag)
Temperature
range
Item
Temperature
Range (°C)
Relative
Humidity Range
Atmospheric
Pressure (mbar)
Operating
15 to 30
10 to 95% non-
condensing
800 to 1,100
Transport
-10 to 70
Up to 95% non-
condensing
500 to 1,100
Storage
-10 to 55
Cooling system
Fan-cooled and thermoelectric cooling (TEC) for laser diode and crystal
Dimensions
Item
Dimensions (W x L x H, mm)
System
370 x 330 x 13
Ophthalmic Table
920 x 630 x 1400
Portable Carrying Case
780 x 530 x 280
Weight
Item
Weight (kg)
System
13
Packed
20
3.1.2
Treatment Laser Specifications
Item
Description
Wavelength
(nm)
532
561
577
670
810
Laser type
OPSL
DPSL
OPSL
Diode
Mode of
operation
Continuous wave
Power
output
2 W maximum to patient’s cornea. Power from
LDU laser aperture displayed on liquid crystal
display (LCD) or 7-inch screen
0.7 W
maximum to
patient’s
cornea.
Power from
LDU laser
aperture
2 W
maximum to
patient’s
cornea via slit
lamp. 3 W
maximum to
patient’s