![Liftkar 045 723 Instruction Manual Download Page 18](http://html1.mh-extra.com/html/liftkar/045-723/045-723_instruction-manual_1911269018.webp)
Liftkar PT I 18
Data for electromagnetic compatibility (EMC)
EMV-general data
Emission
Accordance
RF-Emission according EN
55011
group 1
RF-Emission according EN
55011
class B
Generation of mains
harmonics following IEC
6100-3-2
Battery operation, not applicable
Generation of mains
harmonics following IEC
6100-3-3
Battery operation, not applicable
EMV-table 202
Susceptibility
IEC 60601-test level
actual level
ESD
IEC 61000-4-2
±6kV cd
±8kV ad
±2kV,±4kV ±6kV cd
±8kV ad
Bursts
IEC 61000-4-4
±2kV mains
±1kV I/O
Battery operation, not applicable
Surges
IEC 61000-4-5
±1kV dm
±2kV cm
Battery operation, not applicable
Voltage drops etc.
IEC 61000-4-11
Reduction to
Reduction to
5% for 10ms/ positive amplitude
Battery operation, not applicable
5% for 10ms/ negative amplitude
Battery operation, not applicable
40% for 100ms
Battery operation, not applicable
30% for 500ms
Battery operation, not applicable
0% for 5000ms
Battery operation, not applicable
H-field at 50/60Hz
IEC 61000-4-8
3A/m
No magnetically sensitive sensors or parts,
not applicable
EMV-table 204
Susceptibility
IEC 60601-test level
actual level
Conducted RF
IEC 61000-4-6
3Veff
150kHz to 80 MHz
3Veff
Radiated RF
IEC 61000-4-3
3V/mf
80 MHz to 2,5 GHz
3V/m
EMV-table 206
Output power of transmitter W
Safety distance depending on frequency/m
150kHz to 80MHz 80MHz to 800MHz 800MHz to 2,5 GHz
0,01
0,12m
0,12m
0,24m
0,1
0,37m
0,37m
0,74m
1
1,17m
1,17m
2,34m
10
3,69m
3,69m
7,38m
100
11,67m
11,67m
23,34m
Specifications subject to change without notice.
Summary of Contents for 045 723