2. FEATURES
10
15. Normal Operating Environment:
Overvoltage Category: II, pollution degree:2, altitude:
≤6,562 feet / 2,000m
1. Temperature range: 41
℉
to 104
℉
(5
℃
to 40
℃)
2. Relative humidity:
≤
80%
3. Atmospheric pressure: 86kPa to 106kPa (12.47psi to 15.37psi)
NOTE:
When the storage temperature of the device has been below 41°F (5
℃
)
,
place the device in a location within normal operating temperatures for at
least 4 hours before use.
Operating the equipment outside normal recommended conditions will
reduce its efficiency and lifespan.
16. Electrical category:
II IP21
17. Oxygen Output Temperature:
No higher than +43°F (+6°C) above ambient
temperature.
18. Tube:
Avoid bending the nasal oxygen tube. Tube extensions should not
exceed 49 feet (15.2 meters).
19. Storage and Transport Environment:
1. Temperature Range: 32
℉
to 131
℉
(0
℃
to +55
℃)
2. Relative Humidity Range: 10% to 90%
3. Atmospheric pressure: 70kPa to 106kPa (10.2psi to 15.37psi)
NOTE:
The oxygen concentrator should be stored in an area free from gas and
fumes. During transport, it should always be secured to prevent vibrations
and remain upright at all times.
20.
This device is expected to work for a period of 6 years. For a longer use, please
consult your Equipment Provider and physician.
21.
Software Version: A1.0
22.
Contraindication: Do not use for patients with oxygen allergy or oxygen
poisoning.