121
SPECIFICATIONS
Result Range:
1.1
to
33.3
mmol/L
Calibration: Plasma-equivalent
Sample:
Fresh capillary whole blood
Sample Size:
Minimum
1
microlitre
Test Time:
5
seconds
Assay Method:
Glucose oxidase biosensor
Power Source:
Two AAA alkaline batteries
Battery Life:
About six months at three
tests per day (one test using
backlight)
Glucose Units:
mmol/L
Memory: Over
3
,
000
records
Automatic Shutoff:
Three minutes after inserting a
test strip if sample has not
been applied; one minute after
all other user actions
Size: Approx.
97
mm (L) x
58
mm (W)
x
23
mm (H) [
3.8
in. (L) x
2.3
in. (W) x
0.9
in. (H)]
Weight:
75
grams (with batteries)
Operating Ranges:
Temperature
6–44°
C
Relative humidity
10–90
%
Haematocrit
30–55
%
DCO_06158108A_OUS_OB_GB 5/6/03 1:25 PM Page 121