background image

13

 

C.

 

CONTROL SOLUTION TESTING

How to test with control solution

1.

  Insert a test strip to turn the 

meter on

Make sure the three contact bars are facing 
you and the arrow is pointing down. Push 
the test strip in as far as it will go. 

Do Not

 

bend the test strip.

2.

  Wait for the meter to prompt you to apply a 

control solution sample

After the start-up test screen, your 
last result (blood glucose or control 
solution test) will appear on the 
display, and the back arrow icon  
will blink.

After your last result is displayed, 
your meter will beep and an icon 
appears showing a drop with a test 
strip that fills. This indicates that 
your meter is ready for you to apply 
a control solution sample. You have 
2 minutes to apply a control solution 
sample before the meter turns off.

mg/dL

 

CAUTION: 

Do Not

 swallow control solution; it is not for human 

consumption. 

Do Not

 apply control solution to the skin or 

eyes as it may cause irritation.

When to test with control solution

OneTouch® Select™ Control Solution contains a known amount 
of glucose and is used to check that the meter and the test 
strips are working properly.

Do a control solution test:
•  to practice the test process instead of using blood,
•  whenever you open a new vial of test strips,
•  if you suspect the meter or test strips are not working 

properly,

•  if you have had repeated unexpected blood glucose results 

as described in Applying blood and reading results in 
Section B, or

•  if you drop or damage the meter.

NOTE:

•  Use only OneTouch® Select™ Control Solution with your 

OneTouch® SelectSimple™ Meter.

•  Control solution is available separately. Ask for control 

solution where you get your test strips or contact customer 
service at 1800 225544 or 022-30845544.

•  Control solution tests must be done at room temperature 

20–25°C. Make sure your meter, test strips, and control 
solution are at room temperature before testing.

Summary of Contents for OneTouch SelectSimple

Page 1: ...de Instructions for Use About your System Your OneTouch SelectSimple Meter does not have any buttons Simply insert a test strip to turn the meter on and begin testing Blood Glucose Monitoring System S...

Page 2: ...and detailed information about your system 19 Contact customer service at 1800 225544 or 022 30845544 07 00 AM 11 00 PM or visit http www onetouch co in If you cannot reach Customer Service contact yo...

Page 3: ...240 600 mg dL ACTION PLAN Fill out with your doctor ACTION PLAN Fill out with your doctor SYMPTOMS Increased thirst need to urinate often hunger blurry vision drowsy SYMPTOMS Shaking fast heartbeat sw...

Page 4: ...illary whole blood The OneTouch SelectSimple System is intended for use outside the body in vitro diagnostic use by people with diabetes in a home setting and or by health care professionals in a clin...

Page 5: ...at came with the device Meter Lancing Device Display Test strip port Insert test strip here to turn meter on mg dL is the pre set unit of measure Glucose Range Dots correspond to three pre set glucose...

Page 6: ...appear for 2 seconds All segments of the display should appear briefly on the start up test screen to let you know the meter is working properly mg dL CAUTION If any information is missing from the s...

Page 7: ...a test Use the test strip immediately after removing it from the vial OneTouch Select Test Strips are for single use only Never re use a test strip that had either blood or control solution applied to...

Page 8: ...rotective cover and save it for later use Replace the blue cap by snapping it back on 4 Adjust the depth setting Twist the lancing device cap until the desired setting appears Smaller numbers are for...

Page 9: ...ready for you to apply a blood sample You have 2 minutes to apply a blood sample before the meter turns off 7 Lance your finger Hold the lancing device firmly against the side of your finger Press the...

Page 10: ...tion window to fill completely The blood drop will be drawn into the narrow channel and the confirmation window should fill completely When the confirmation window is full this means you have applied...

Page 11: ...information on LOW HIGH and VERY HIGH results Example 1 Knowing when your test result is LOW HIGH or VERY HIGH Your meter has two colour dots below the display Each dot corresponds to a pre set LOW H...

Page 12: ...r proper disposal CAUTION If your blood glucose is high and you test when the temperature is at the low end of the operating range 10 C the result on your meter may be lower than your actual glucose l...

Page 13: ...severely dehydrated If you think you are severely dehydrated contact your healthcare professional immediately High glucose results If your test result is higher than 180 mg dL it may mean hyperglycae...

Page 14: ...physical exercise or you think your glucose level is changing rapidly You are ill or during times of stress You are concerned about the possibility of hypoglycaemia or an insulin reaction such as whe...

Page 15: ...22 30845544 1 Insert a sterile lancet and snap on the clear cap 2 Adjust the depth setting You may need a deeper setting to get a sufficient sample for testing Twist the clear cap toward the larger nu...

Page 16: ...ar the blood sample If the drop of blood runs or spreads due to contact with hair or a line in your palm do not use the sample Try lancing again in a smoother area If bruising occurs at an alternate s...

Page 17: ...r human consumption Do Not apply control solution to the skin or eyes as it may cause irritation When to test with control solution OneTouch Select Control Solution contains a known amount of glucose...

Page 18: ...g with the unit of measure NOTE If the HIGH LOW alarm is activated during control solution testing do not be concerned or take any action The HIGH LOW alarm should not be used to evaluate control solu...

Page 19: ...u used If the result you get is not within the appropriate range the meter and test strips may not be working properly Repeat the control solution test 7 Cleaning Clean the top of the control solution...

Page 20: ...tests but you should replace the battery as soon as possible The icon will continue to blink on all other screens until you replace the battery A blinking empty battery icon will appear by itself on t...

Page 21: ...neTouch SelectSimple Blood Glucose Monitoring System does not need any special maintenance Storing your system Store your meter test strips and control solution in your case after each use Store each...

Page 22: ...ampened with this solution Immerse the caps only in this solution for 30 minutes After disinfecting rinse briefly with water and allow both to air dry If your lancing device is different than the one...

Page 23: ...h the meter What to do Repeat the test with a new test strip see Section B Testing your blood glucose If this message continues to appear contact customer service at 1800 225544 or 022 30845544 What i...

Page 24: ...ange 10 44 C and repeat test with a new test strip see Section B Testing your blood glucose If this error message appears again contact customer service at 1800 225544 or 022 30845544 Error 4 continue...

Page 25: ...recommendations Although this message could be due to a test error it is safer to treat first and then do another test NOTE There is no Error 6 message What it means You may have a very high blood gl...

Page 26: ...have installed a new battery OR Your meter was unable to recall the last result What to do You can still perform a blood glucose test and get an accurate test result Contact customer service at 1800 2...

Page 27: ...other type of meter one that provides whole blood calibrated results you may notice that your test results with the OneTouch SelectSimple Meter are approximately 12 higher OneTouch SelectSimple Meter...

Page 28: ...le Meter will be free of defects in material and workmanship for three years valid from the date of purchase The guarantee extends only to the original purchaser and is not transferable Technical spec...

Page 29: ...ter results that match the laboratory test Therefore 99 7 of the total results obtained with the OneTouch SelectSimple System achieved the goal suggested by the diabetes experts System Accuracy Result...

Page 30: ...g dL Std Error Sy x mg dL R2 0 951 to 0 967 2 488 to 0 917 11 054 0 989 Precision Within Run Precision Repeatablity 100 Venous Blood Tests Target Glucose mg dL Mean Glucose mg dL Standard Deviation mg...