MESSAGE
PROBLEM
WHAT TO DO
1) The blood sample
Repeat the test with a
was too small or
new Test Strip and a
smeared.
large, shiny drop of
blood.
2) The Test Strip was not
Repeat the test. This
inserted all the way
time push the Test Strip
into the Test Strip
all the way into the
Holder.
Test Strip Holder.
Your Meter may have an
Call LifeScan Technical
electronic problem.
Services.
The Check Strip test result is
Clean the Meter and
outside the acceptable range.
Check Strip. Repeat the
Check Strip test.
A 105
(example)
Anot ok
redo A
not ok
not
enough
blood
retest
50