20
Lifeloc FC10
Plus
www.Lifeloc.com
|
303-431-9500
A Calibration Check
simply verifi es the FC10
Plus
was calibrated correctly and is within the
acceptable accuracy range.
Calibration check is also referred to as “External Calibration
Check,” “Accuracy Check,” “Verifi cation” and “Cal Check.”
Lifeloc recommends you perform a calibration check on your FC10
Plus
:
Once
every 30 days
Or
, at intervals specifi ed by your Internal Policies, Quality Assurance Plan, or
State
Regulations.
Selecting the Standard Type
Press the
Function
button until the display
reads “CALIBRATION”.
Press the
Execute
button. The display reads
either “WET CHECK” or “DRY CHECK.”
Press the
Function
button until the display
reads “CAL SETTINGS”.
Press the
Execute
button. The display reads
“CAL STANDARD”.
Press the
Function
button until the display
reads “STANDARD TYPE”.
Press the
Execute
button to choose between
“DRY GAS” and “WET BATH”.
Press the
Function
button to save settings. Momentarily press
Power
to return to the
testing mode.
Standard Type
Wet Bath
Change
Calibration/Cal Check Lock-out Explained
The calibration/cal check lock-out allows you to prevent usage of your FC10
Plus
if it is not
calibrated or cal checked in a specifi ed time period. Time can be registered as days or
number of tests.
Beginning 48 hours before the specifi ed lock-out time, the unit will display “WARNING
CAL EXPIRING,” “WARNING CHECK EXPIRING” or both.
When the time period has elapsed, the unit will display “CAL EXPIRED,” “CHECK
EXPIRED” or both and not allow testing until it is calibrated and/or cal checked.
You can specify time periods up to 999 days or 9999 tests between calibrations or cal
checks. You can also disable the calibration/cal check lock-out feature. (See p.21)
Calibration Settings