background image

P O W E R I N G   T H E   L I F E C H O I C E

The LifeChoice POC can be powered in four different ways – the internal battery, an AC power supply,
DC power supply or the external battery. The rechargeable batteries are maintenance-free and can
only be exchanged at the factory.

1.

Internal  Battery:

A  rechargeable  internal

battery  is  located  within  each  LifeChoice
POC unit. When it is fully charged, it supplies
power to the Life Choice POC for up to 2
hours. When using internal battery power,
the green LED light above the 

Battery Check

will illuminate and flash slowly.

When the AC, DC or external battery is plugged into the LifeChoice POC, whether powered
on or off, the green LED light above the 

Battery Check

will illuminate solid green. This is

verification that the external source has been recognized. When using an external source to
power the LifeChoice POC, verify that the green LED light above the 

Battery Check

is solid

green indicating the system is being powered by the external source. If not connected to an
AC or DC adapter, the internal battery will slowly discharge over time even when not in use.
Always check the battery level prior to use to ensure adequate charge level.

2. 

AC Power Supply:

An AC power supply

allows the LifeChoice POC to be connected
to a 100-240 volt, 50/60 Hz outlet. The
power supply converts 100-240 volt AC to a
DC  voltage  for  the  LifeChoice  POC  unit.
When the unit is operated with the AC power
supply, power from the AC outlet powers the
unit and recharges the internal and/or the
external battery simultaneously. 

When  using  an  external  source  to  power
the LifeChoice POC, verify that the green LED light above the 

Battery Check

is solid green

indicating the system is being powered by the external source.

3.

DC Power Supply:

A DC power supply can

be connected from the LifeChoice POC unit
to an automobile’s 

(boat, motor home, etc.)

12-volt

DC outlet. When the system is connected
to  the  automobile’s/vehicle’s  DC  outlet,
power  from  the  automobile/vehicle
battery powers the LifeChoice POC unit
and recharges the internal and/or external
battery simultaneously.

When using an external source to power the
LifeChoice POC, verify that the green LED
light  above  the 

Battery  Check 

is  solid 

7

L i f e C h o i c e   P o r t a b l e   O x y g e n   C o n c e n t r a t o r

Summary of Contents for XYC276

Page 1: ...Operator s Manual...

Page 2: ...YC276 REV J Oct 2012 i 3500 Comsouth Drive Suite 100 Austin TX 78744 USA www InovaLabs com Phone 1 800 220 0977 0459 Represented in Europe by QNET BV Hommerterweg 286 6436 AM Amstenrade The Netherland...

Page 3: ...hoice POC Mobility Package 5 LifeChoice POC Part Names 5 Operating Instructions 6 Powering the LifeChoice POC 7 Resetting the External Battery 9 Battery Charging 9 Operating Procedure 10 Alarms Light...

Page 4: ...LifeChoice Oxygen Concentrator is used on a prescriptive basis by patients who are diagnosed as requiring supplemental oxygen This oxygen concentrator will provide supplemental high concentration oxy...

Page 5: ...physician or medical equipment supplier for the type of back up system required 9 The LifeChoice POC is designed to be used in the provided approved carry case 10 The LifeChoice should be located so...

Page 6: ...equivalent must be used with the LifeChoice POC to provide oxygen from the concentrator A standard single lumen type cannula 7 foot 2 meter is recommended for use An Adult Nasal Cannula like 1107 by V...

Page 7: ...nasal cannula 1 2 3 L I F E C H O I C E P O C M O B I L I T Y P A C K A G E X Y C 1 0 0 The LifeChoice POC can be operated using AC power DC power or battery power during transport while inside the a...

Page 8: ...source is notrecognized Ifchargingtheunitthroughthe externalbatterywhenthisoccurs thechargewill not reach the internal battery This may be an indicationtheexternalbatteryneedstobereset Instructions fo...

Page 9: ...ry will slowly discharge over time even when not in use Always check the battery level prior to use to ensure adequate charge level 2 AC Power Supply An AC power supply allows the LifeChoice POC to be...

Page 10: ...y is required for extended time away from an AC or DC power source it is recommended to power the LifeChoice POC with the external battery first and then the internal battery for maximum combined batt...

Page 11: ...ay be illuminated but the actual charge could be between 85 100 The internal battery cannot be overcharged and it is recommended to leave the LifeChoice POC plugged in via the AC or DC adapter as ofte...

Page 12: ...T I N G P R O C E D U R E 1 Locate and position the LifeChoice POC so that the intake vents and exhaust fan vents are not obstructed The unit particularly the air intake vents and the exhaust fan vent...

Page 13: ...s Power once to turn on An audible beep will occur and the green LED light above the Power Active High Mode and Pulse Setting of 1 will illuminate This indicates that LifeChoice POC is powered on is r...

Page 14: ...Choice POC for the first time or upon subsequent use To accurately measure the internal battery charge the unit must be powered on and disconnected from the AC DC power supply or external battery The...

Page 15: ...wer outlet or to an AC power outlet the unit operates while recharging the internal battery simultaneously The level of battery charge is indicated by disconnecting the LifeChoice POC from the AC or D...

Page 16: ...has detected an inhalation breath and is delivering a pulse of oxygen LifeChoice POC powered by internal battery No LifeChoice POC is functioning normally LifeChoice POC is functioning normally No Li...

Page 17: ...es of battery power Critical Low Battery Alarm Approximately 10 seconds of battery power Fast audible beep Unit shuts down Connect the LifeChoice POC unit to a DC adapter an AC adapter or change to an...

Page 18: ...an and remove any debris from the intake filter mesh Do not poke or gouge the mesh filter just gently rub the cotton swab cleaner along the intake vent areas The LifeChoice POC does not need to be rem...

Page 19: ...6A External AC Power Adapter XYC103AC Input 100 240V 2 5A Output 19V 6 3A External DC Power Adapter XYC104DC Input 11 5 16VDC 5 6A Output 19V 6 3A External Rechargeable Battery XYC102 Input 14 2 23 7...

Page 20: ...quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV line to line 2kV line to earth 1 kV line to line 2kV line to earth Mains power quality should be that of...

Page 21: ...nt should be used no closer to any part of the Model XYC100 including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recomm...

Page 22: ...fic and medical bands between 150 kHz and 80 MHz are 6 765 MHz to 6 795 MHz 13 553 MHz to 13 567 MHz 26 957 MHz to 27 283 MHz and 40 66 MHz to 40 70 MHz NOTE 3 An additional factor of 10 3 has been in...

Page 23: ...d warranties must be purchased at the time of original purchase of the LifeChoice The Limited Warranty Excludes From Coverage 1 Defects caused by or related to a Abuse misuse negligence or accident b...

Page 24: ...INDIRECT EXEMPLARY OR PUNITIVE DAMAGES REGARDLESS OF THE FORM OF THE CLAIM WHETHER IN CONTRACT TORT OR OTHERWISE ARISING OUT OF OR RELATING TO INOVA LABS PRODUCTS AND EVEN IF INOVA LABS HAS BEEN ADVI...

Page 25: ...arry case Included with purchase of LifeChoice POC Extra small waist belt strap for external battery pack Included with purchase of LifeChoice POC Product Product Number Description 23 L i f e C h o i...

Page 26: ...and nasal cannula Each LifeChoice POC comes with a limited three 3 year warranty Model Number XYC200 Includes unit 7 5 w x 3 125 d x 9 5 h 4 9 pounds 4 way carry case which can be used as a convertib...

Page 27: ...e The LifeChoice POC device contains lithium ion batteries and the end users should contact Inova Labs or their local distributor for disposal collection and recycling options and terms and conditions...

Page 28: ...vaLabs com Manufactured and Distributed by Inova Labs Inc 3500 Comsouth Drive Suite 100 Austin Texas 78744 Corporate Office 1 512 617 1700 2012 Inova Labs Inc All rights reserved Assembled in the USA...

Reviews: