MAINTENANCE

AIRFILTER: The airfilteron the top panel of the compressor unit is bright white when new. With use, it will turn grey, which 
means it is time to replace it. It must be replaced immediately if it gets wet or becomes clogged. The air filter lasts up to 500 
hours of use before replacement. A general rule is to replace the air filter whenever the nebulizer unit is replaced (every 6 
months). To replace the air filter, ensure the compressor is turned off, open the air filter cover to take out the air filter and 
replace it with a new one. Place the cap back into the socket of the compressor and snap the filter caver back in place.
NOTE
Do not attempt to wash and reuse tha air filter. Do not use substitute filter material. Compressor damage may result.

AIR FILTER COVER

AIR FILTER

Dimensions (Compressor 
only):

27.4x16.7x10.6

Weight (Compressor only):

4. 1 pound(1.86kg)

Electrical Requirements:

230V

Power Consumption:

68W

High Hydrostatical pressure:

860-1060HPA

Humidity:

30%-75%

Operation cycles:

Operation for 20 min.
0fffor40 min.

NEBULIZER SPECIFICATIONS