Overview

The EVOS

®

 XL Core Imaging System

 

has two 

types of controls: manual and onscreen. Manual 
controls include the illumination wheel, focus 
knobs, freeze and Save buttons, objective turret, 
and the optional mechanical stage X-Y axis 
knobs. Onscreen controls appear at the bottom 
of the screen when the mouse rolls over the 
controls area. using a mouse is optional; you 
can operate the eVOS

®

 XL core imaging System 

using only manual controls. 

Note: The EVOS

®

 XL Core Imaging System User 

Guide is provided on the USB flash drive. It is also 
available for downloading from the EVOS

®

 XL Core 

Imaging System product page at   
www.lifetechnologies.com.

Operating Environment

 

Place the eVOS

®

 XL core imaging System 

on a level surface away from vibrations from 
other pieces of equipment.

 

Allow at least 5 cm (2 in) of free space at 
the back of the LcD housing to allow for 
proper ventilation and prevent overheating 
of electronic components.

 

Set up the instrument away from direct light 
sources, such as windows. Ambient room 
lighting can enter the imaging path and 
affect the image. 

 

Operating temperature range: 4°–32°c 
(40°–90°f).

 

Relative humidity range: 30–90%.

 

2

Onscreen  

Controls

Setup

1.

 

connect the power adapter to the power 
jack 

 on the back of the instrument base 

and plug the cord into an outlet.

2.

  Plug the wireless mouse receiver into 

either of the USB ports 

 located on the 

right side of the LcD screen. install the 
battery and switch on the mouse.

3.

  Optional: if the instrument is equipped with 

a mechanical stage, install the mechanical 
stage following the instructions provided in 
the user guide.

4.

 

unpack the 

glass stage plate

 

 and insert 

it into the stage opening (if a mechanical 
stage is installed, move it all the way back to 
expose the stage opening).

5.

 

Switch the instrument On using the 
power switch 

 on the right side of the 

instrument. Startup takes approximately 
one minute. if the display remains black 
after startup, adjust the illumination.

Note: For detailed setup instructions, refer 
to the EVOS

®

 XL Core Imaging System User 

Guide provided on the USB flash drive.

Basic Operation

1.

 

Turn on the instrument using the power 
switch
 

 on the right side of the base.

2.

  Plug a uSB flash drive into one of the USB 

ports 

 on the right side of the instrument.

3.

  Place the sample on the glass stage 

plate 

. if the instrument is equipped with 

a mechanical stage, secure a vessel with 

the stage clip 

 to hold the sample over 

the stage plate

.

Note: Place slides with coverslips face up.

4.

  Set the objective turret 

 to the desired 

magnification. 

5.

  Turn the phase turret 

 to the position 

that corresponds to the selected objective 
and contrast method.

6.

  Bring the sample into focus using the 

coaxial focus knob 

.

7.

  Adjust the illumination intensity if necessary, 

using the onscreen illumination slider 

 

or the illumination wheel 

8.

  use the onscreen Freeze/Live button 

11

 

or the manual Freeze button 

12

 to acquire 

the image. 

9.

  use the onscreen Save button 

13

 or the 

manual Save button 

14

 to save the image.

10.

 click the onscreen Freeze/Live button 

11

 

or the manual Freeze button 

12

 to return 

to live viewing.

Note: The Freeze/Live button toggles 
between Freeze and Live.

3

4

Note: Onscreen controls are available only when 

mouse is installed. Roll over bottom of screen with 

mouse to activate controls. 

14

12

Dimensions

The 

eVOS

®

 XL core imaging System will fit in cell 

culture hoods that are at least 18 inches (457 mm) 
deep. if your cell culture hood is smaller, it may be 
necessary to turn the instrument at a slight angle 
to fit.

DEPTH

16.0 in (40.6 cm)

WIDTH

12.5 in (31.8 cm)

HEIGHT, TRANSPORT

13.5 in (34.3 cm)

HEIGHT, DISPLAY

21.0 in (55.3 cm)

13

11

Summary of Contents for EVOS XL Core

Page 1: ...rable right to use the purchased amount of the product only to perform internal research for the sole benefit of the purchaser No right to resell this product or any of its components is conveyed expressly by implication or by estoppel This product is for internal research purposes only and is not for use in commercial applications of any kind including without limitation quality control and comme...

Page 2: ...he way back to expose the stage opening 5 Switch the instrument ON using the power switch on the right side of the instrument Startup takes approximately one minute If the display remains black after startup adjust the illumination Note For detailed setup instructions refer to the EVOS XL Core Imaging System User Guide provided on the USB flash drive Basic Operation 1 Turn on the instrument using ...

Page 3: ...he way back to expose the stage opening 5 Switch the instrument ON using the power switch on the right side of the instrument Startup takes approximately one minute If the display remains black after startup adjust the illumination Note For detailed setup instructions refer to the EVOS XL Core Imaging System User Guide provided on the USB flash drive Basic Operation 1 Turn on the instrument using ...

Page 4: ...he way back to expose the stage opening 5 Switch the instrument ON using the power switch on the right side of the instrument Startup takes approximately one minute If the display remains black after startup adjust the illumination Note For detailed setup instructions refer to the EVOS XL Core Imaging System User Guide provided on the USB flash drive Basic Operation 1 Turn on the instrument using ...

Page 5: ...rable right to use the purchased amount of the product only to perform internal research for the sole benefit of the purchaser No right to resell this product or any of its components is conveyed expressly by implication or by estoppel This product is for internal research purposes only and is not for use in commercial applications of any kind including without limitation quality control and comme...

Page 6: ...rable right to use the purchased amount of the product only to perform internal research for the sole benefit of the purchaser No right to resell this product or any of its components is conveyed expressly by implication or by estoppel This product is for internal research purposes only and is not for use in commercial applications of any kind including without limitation quality control and comme...

Reviews: