Powering the LI-830 and LI-850
A power supply—either a battery or the universal power adapter—can be connected
to the LI-830 and LI-850 power jack or to pins 1 and 2 on the terminal strip. The
instruments require 12 to 30 VDC. The power supply must be able to provide 14
watts (1.2 A at 12 V) during warmup and about 3.6 watts (0.3 A at 12 V) during
normal operation.
Using the universal power adapter
The included power supply is compatible with mains power in most localities glob-
ally (input 100 to 240 VAC, 50 to 60 Hz; output 12 VDC, 2.5 amps). You may need
to install different blades from the outlet adapter kit to fit your wall outlets.
PU
SH
OPEN
LOCK
N. America, Japan
United Kingdom
Australia
Europe
Turn clockwise
to lock
Push button, then
turn counterclock-
wise to release
Outlet Adapter Kit
PU
S
H
OPEN
LOC
K
PU
S
H
O
PEN
LOC
K
PU
S
H
OPEN
L
O
CK
PU
S
H
L
OCK
OPEN
Using the terminal strip power connectors
The terminal strip has connectors for any power supply that has bare leads. Pin 1 is
the positive (+) terminal, and pin 2 is the negative (-) terminal.
Section 1. General description
1-7
Powering the LI-830 and LI-850