Appendix A. Measurement overview
To measure all three gases, the laser frequency scans across the full spectral range of
interest in about 0.25 seconds (see
below). It uses the reliably fixed grid
spacing in combination with spectroscopic data to determine the exact frequency of
each mode. The instrument uses two ring down events—one at the beginning and
one at the end of a scan—to normalize the measured signal.
Absorption is computed at each resonance mode by comparing the normalized
sample photodiode signal with the reference photodiode signal. The concentration
of each gas is then determined by using a fitting algorithm that compares the meas-
ured absorption spectrum with internally stored high resolution spectrum.
Figure A-2. Example of a scan showing the absorption features for a number of gases. The
upper line (dark blue) is the reference signal, and the lower line (green) is the sample signal
at each of cavity resonance modes.
A-2
LI-7815 CO₂/H₂O Gas Analyzer
Summary of Contents for LI-7815
Page 1: ...LI 7815 CO2 H2O Trace Gas Analyzer Instruction Manual...
Page 2: ......
Page 10: ...viii...
Page 30: ...Section 2 Getting started 2 14 LI 7815 CO H O Gas Analyzer...
Page 38: ...Section 3 Applications of the LI 7815 3 8 LI 7815 CO H O Gas Analyzer...
Page 42: ...Section 4 Software overview 4 4 LI 7815 CO H O Gas Analyzer...
Page 54: ...Section 4 Software overview 4 16 LI 7815 CO H O Gas Analyzer...
Page 62: ...Section 5 Troubleshooting 5 8 LI 7815 CO H O Gas Analyzer...
Page 70: ...Section 6 Maintenance 6 8 LI 7815 CO H O Gas Analyzer...
Page 88: ...Appendix B Additional safety information B 6 LI 7815 CO H O Gas Analyzer...
Page 92: ...Appendix C Specifications C 4 LI 7815 CO H O Gas Analyzer...
Page 97: ......