tubes. Additional nuts and ferrule sets are include in the accessories kit.
Warning:
Do not exceed the maximum absolute pressure rating of 110 kPa.
Advertissement:
Ne dépassez pas la pression nominale absolue maximale de 110 kPa.
Air outlet
The instrument exhausts sample air through the outlet. Do not substantially restrict
or apply a vacuum to the outlet.
Rechargeable batteries
Part Number
442-11807
Two batteries are provided with the instrument. We recommend leaving at least one
battery in the instrument when you are using it. The batteries are 14.4 V lithium ion
battery packs with 6.8 amp-hour capacity. Two batteries can provide up to 8 hours of
operation under most conditions. Batteries are shipped in a partially discharged
state. They should be fully charged prior to the first use. Storing the batteries in a dis-
charged state can result in damage to the batteries. Batteries that are installed in the
instrument will be charged when the instrument is powered though the universal
power adapter.
Warning:
Lithium ion batteries may cause environmental damage or damage to human
health if they are disposed of improperly. Do not dispose of the batteries in unsorted
municipal waste or an incinerator. Many localities have battery recycling facilities that
will accept lithium ion batteries. Check your local regulations to determine how to safely
dispose of the batteries.
Advertissment:
Les batteries lithium ion peuvent causer des dommages à
l'environnement ou endommager la santé humaine si l’on se débarrasse d’elles de man-
ière inappropriée. Ne vous débarrassez pas des batteries dans des déchets municipaux
non triés ou dans un incinérateur. Plusieurs localités ont des installations de recyclage
de batteries qui accepteront les batteries lithium ion. Vérifiez vos réglementations loc-
ales pour déterminer comment se débarrasser des batteries en toute sécurité.
Universal power adapter
Part Number
9968-232
The universal power adapter is an AC-to-DC adapter that powers the instrument and
charges the batteries when it is plugged into an AC power outlet. The power adapter
provides 24 VDC (9.17 amps maximum; 220 watts maximum). It is compatible with
Section 1. Introduction to the LI-7815
1-3
Rechargeable batteries
Summary of Contents for LI-7815
Page 1: ...LI 7815 CO2 H2O Trace Gas Analyzer Instruction Manual...
Page 2: ......
Page 10: ...viii...
Page 30: ...Section 2 Getting started 2 14 LI 7815 CO H O Gas Analyzer...
Page 38: ...Section 3 Applications of the LI 7815 3 8 LI 7815 CO H O Gas Analyzer...
Page 42: ...Section 4 Software overview 4 4 LI 7815 CO H O Gas Analyzer...
Page 54: ...Section 4 Software overview 4 16 LI 7815 CO H O Gas Analyzer...
Page 62: ...Section 5 Troubleshooting 5 8 LI 7815 CO H O Gas Analyzer...
Page 70: ...Section 6 Maintenance 6 8 LI 7815 CO H O Gas Analyzer...
Page 88: ...Appendix B Additional safety information B 6 LI 7815 CO H O Gas Analyzer...
Page 92: ...Appendix C Specifications C 4 LI 7815 CO H O Gas Analyzer...
Page 97: ......