Guided Tours
Tour #3: Controlling Chamber Conditions
3-34
Using the LI-6400
3
Experiment #6
Dynamic Response of the CO
2
Mixer
Make strip charts for this using the Real Time Graphics editor (press
4
, then
F4
) to define two curves as shown in Figure 3-25. (Hint:
CO2R
and
CO2S
have ID numbers -1 and -2 in the list of variables that appears when you edit
Y axes.) Save the configuration as ÒCO2Ref&SampleÓ, or something similar.
1
Close the empty chamber
ItÕs another no-leaf experiment.
2
Fixed flow at 500 µmol s
-1
Press
2
f2
F
500
enter
.
3
Set the mixer for a 400 µmol mol
-1
reference target
Press
f3
R
400
enter
.
4
Turn on Real Time Graphics
Press
4
then
f1
to start the graph you should have defined in Figure 3-25.
5
After 1 minute, change reference target to 900 µmol mol
-1
.
When the
CO2S
curve flattens out, press
escape
2
f3
900
enter
.
Points to Remember
¥ Soda lime must remain on full scrub when using the mixer
¥ The lowest stable value is typically between 30 and 50 µmol mol
-1
.
¥ Mixer adjusts quicker when lowering the concentration.
Real Time Graphics
__X_AXIS_________ __Y_AXIS___________
1:Æt=180 CO2R:0 to 1000
2:Æt=180 CO2S:0 to 1000
Edit #Curvs Help Cancel OK
2
+
Figure 3-25. Strip chart definition for Experiment #6.
120 second
strip charts
Sample and ref-
erence CO2 use
fixed axes
Summary of Contents for LI-6400
Page 1: ...Using the LI 6400 Portable Photosynthesis System ...
Page 15: ...Part I The Basics ...
Page 16: ......
Page 174: ...Making Measurements Answers to Questions 4 56 Using the LI 6400 4 ...
Page 175: ...Part II Useful Details ...
Page 176: ......
Page 200: ...Standard Tools Power ON Hooks 5 24 Using the LI 6400 5 ...
Page 214: ...Real Time Data Real Time Graphics 6 14 Using the LI 6400 6 ...
Page 234: ...Environmental Control Light Control 7 20 Using the LI 6400 7 ...
Page 244: ...Light Sensor Considerations Gallium Arsenide Phosphide GaAsP Sensor 8 10 Using the LI 6400 8 ...
Page 288: ...Data Logging Making Your Own AutoPrograms 9 44 Using the LI 6400 9 ...
Page 289: ...Part III Working With Files ...
Page 290: ......
Page 312: ...The LPL File System Troubleshooting 10 22 Using the LI 6400 10 ...
Page 340: ...Downloading Files Using a Data Capture Program 11 28 Using the LI 6400 11 ...
Page 375: ...Part IV Configuration Issues ...
Page 376: ......
Page 420: ...Defining User Variables Old Style vs New Style 15 18 Using the LI 6400 15 ...
Page 454: ...Using an Energy Balance Further Reading 17 12 Using the LI 6400 17 ...
Page 455: ...Part V Maintenance Troubleshooting ...
Page 456: ......
Page 572: ...Troubleshooting Useful Information 20 46 Using the LI 6400 20 ...
Page 593: ...Part VI Programming ...
Page 594: ......
Page 622: ...Programming with LPL Compiler Directives 22 28 Using the LI 6400 22 ...
Page 846: ...Index I 16 Using the LI 6400 ...