Data Logging
AutoProgram Descriptions
9-30
Using the LI-6400
9
Note: Bug in 3.01
ÒRemote ControlÓ in OPEN 3.01 does not function correctly. This has been
fixed in OPEN 3.2.
ÒTimedLampÓ
TimedLamp
allows you to program timed changes in the 6400-02 or -02B
LED source. For example, you may want to set the light level to
200 µmol m
-2
s
-1
for several minutes, then jump to 2000 µmol m
-2
s
-1
for
20 seconds, then drop back to the original value for several more minutes.
TimedLamp
lets you do this. You can also specify the logging frequency for
each lamp level. You may, for example, want to record data less frequently
during the initial stability period, and very frequently when the light changes.
■
■
■
■
TimedLamp
prompts
1
Get Data from Keyboard or File
Press
K
to enter the programÕs set points from the keyboard (if you do this, go
F
to use a previously entered set of points (if you do this,
go to Step 2).
2
Select TimedLampÕs Steps
You are asked to select a file from those in the directory Ò
/User/Configs/Auto-
Progs/TimedLamp Defaults
Ó.
Get data...
K - from Keyboard
F - from File
Timed Lamp Program
DIR: /User/Configs/AutoProgs/TimedLamp Def
"JoeÕs test"
"Dflts"
Select TimedLamp Specs
+
"Sunfleck 4"
"Sunfleck 2"
Dir
¨
¨
¨
CANCEL
¨
SELECT
Summary of Contents for LI-6400
Page 1: ...Using the LI 6400 Portable Photosynthesis System ...
Page 15: ...Part I The Basics ...
Page 16: ......
Page 174: ...Making Measurements Answers to Questions 4 56 Using the LI 6400 4 ...
Page 175: ...Part II Useful Details ...
Page 176: ......
Page 200: ...Standard Tools Power ON Hooks 5 24 Using the LI 6400 5 ...
Page 214: ...Real Time Data Real Time Graphics 6 14 Using the LI 6400 6 ...
Page 234: ...Environmental Control Light Control 7 20 Using the LI 6400 7 ...
Page 244: ...Light Sensor Considerations Gallium Arsenide Phosphide GaAsP Sensor 8 10 Using the LI 6400 8 ...
Page 288: ...Data Logging Making Your Own AutoPrograms 9 44 Using the LI 6400 9 ...
Page 289: ...Part III Working With Files ...
Page 290: ......
Page 312: ...The LPL File System Troubleshooting 10 22 Using the LI 6400 10 ...
Page 340: ...Downloading Files Using a Data Capture Program 11 28 Using the LI 6400 11 ...
Page 375: ...Part IV Configuration Issues ...
Page 376: ......
Page 420: ...Defining User Variables Old Style vs New Style 15 18 Using the LI 6400 15 ...
Page 454: ...Using an Energy Balance Further Reading 17 12 Using the LI 6400 17 ...
Page 455: ...Part V Maintenance Troubleshooting ...
Page 456: ......
Page 572: ...Troubleshooting Useful Information 20 46 Using the LI 6400 20 ...
Page 593: ...Part VI Programming ...
Page 594: ......
Page 622: ...Programming with LPL Compiler Directives 22 28 Using the LI 6400 22 ...
Page 846: ...Index I 16 Using the LI 6400 ...