Making Measurements
Matching the Analyzers
4-38
Using the LI-6400
4
Ò
CO2S Has Changed
Ó
The sample cell CO
2
concentration at the start of match mode is retained, and
periodically compared to subsequent values as a stability check. Whenever
the difference exceeds 3.0 µmol mol
-1
, a warning appears (Figure 4-7).
Since the sample CO
2
concentration should not be affected by match mode,
drift in this value indicates a problem. The cause for such drift is either unsta-
ble incoming CO
2
, a leak, or a sudden change in photosynthetic rate. While
in match mode, if the sample CO
2
is stable but the reference varies, then there
is a leak in the chamber exhaust tube, or else a problem with the match valve
itself.See page 20-21 for troubleshooting help.
The best way to prevent this message is to not enter match mode when the
sample cell concentrations are unstable.
Ò
Excessive Deltas
Ó
If
MATCH
is pressed, and the difference between sample and reference IR-
GAs exceeds 10 µmol mol
-1
for CO
2
or 1 mmol mol
-1
for H
2
O, a warning
will be displayed:
IRGA Match Mode
ÒCONFIG.JoeÓ
CO2R_µml CO2S_µml H2OR_mml H2OS_mml
330.1 329.2 18.24 18.55
Prev: 00:22:01 CO2=402 H2O=18.6
exit
MATCH
CO2S has changed 4.5 ppm (leak?)
Message Line
IRGAs
Figure 4-7. The leak message appears if the CO2S_µml value is more than
3 µmol mol
-1
from the value it had when match mode was entered.
WARNING
Excessive Deltas!
-Chamber closed?
-Match valve working?
Still want to match?
Press (Y/N)
Summary of Contents for LI-6400
Page 1: ...Using the LI 6400 Portable Photosynthesis System ...
Page 15: ...Part I The Basics ...
Page 16: ......
Page 174: ...Making Measurements Answers to Questions 4 56 Using the LI 6400 4 ...
Page 175: ...Part II Useful Details ...
Page 176: ......
Page 200: ...Standard Tools Power ON Hooks 5 24 Using the LI 6400 5 ...
Page 214: ...Real Time Data Real Time Graphics 6 14 Using the LI 6400 6 ...
Page 234: ...Environmental Control Light Control 7 20 Using the LI 6400 7 ...
Page 244: ...Light Sensor Considerations Gallium Arsenide Phosphide GaAsP Sensor 8 10 Using the LI 6400 8 ...
Page 288: ...Data Logging Making Your Own AutoPrograms 9 44 Using the LI 6400 9 ...
Page 289: ...Part III Working With Files ...
Page 290: ......
Page 312: ...The LPL File System Troubleshooting 10 22 Using the LI 6400 10 ...
Page 340: ...Downloading Files Using a Data Capture Program 11 28 Using the LI 6400 11 ...
Page 375: ...Part IV Configuration Issues ...
Page 376: ......
Page 420: ...Defining User Variables Old Style vs New Style 15 18 Using the LI 6400 15 ...
Page 454: ...Using an Energy Balance Further Reading 17 12 Using the LI 6400 17 ...
Page 455: ...Part V Maintenance Troubleshooting ...
Page 456: ......
Page 572: ...Troubleshooting Useful Information 20 46 Using the LI 6400 20 ...
Page 593: ...Part VI Programming ...
Page 594: ......
Page 622: ...Programming with LPL Compiler Directives 22 28 Using the LI 6400 22 ...
Page 846: ...Index I 16 Using the LI 6400 ...