![LI-COR LI-600 Instruction Manual Download Page 91](http://html.mh-extra.com/html/li-cor/li-600/li-600_instruction-manual_1906650091.webp)
LI-600 Porometer/Fluorometer Instruction Manual
9-4
where (
g
sw
) is the one-sided stomatal conductance to
water vapor, and (
g
bw
) is the one-sided boundary layer
conductance, both in (mol H
2
O m
-2
s
-1
).
Boundary layer conductance in the LI-600 cuvette is a
function of the flow rate (equation 9-10).
The actual measurements for water vapor being made in
the LI-600 are done using relative humidity sensors on
the inlet and outlets of the cuvette. In order to calculate
the mass balance as described above, the relative humid-
ities must be converted to molar fractions, requiring
additional measurements for the temperature of the
sensors (
T
ref
) and the atmospheric pressure (
Press
).
The vapor pressures are calculated from the relative
humidity measurement and a saturation vapor function
from Buck (1981):
9-7
where
satvap(T)
is the saturation vapor pressure (kPa) at
the given temperature T (°C)
9-8
where
VP
x
is the calculated vapor pressure (kPa) with
x
is 'cham','ref' or 'leaf',
rh
x
is the relative humidity (%)
measured by the LI-600 sensors and
T
ref
(°C) is the tem-
perature read by a thermistor in the block adjacent to
the RH sensors. For calculation of leaf vapor pressure, it
is assumed that the internal air-space is at saturation,
i.e., relative humidity within the leaf is 100%, which
reduces the equation to equation 9-7.
Summary of Contents for LI-600
Page 1: ...Using the Ll 600 Porometer Fluorometer Instruction Manual ...
Page 41: ...4 14 ...
Page 69: ...7 8 ...
Page 117: ...10 4 ...
Page 119: ...A 2 ...
Page 123: ...B 4 ...
Page 130: ......