LI-COR LI-3000C Battery Replacement Instructions Download Page 3

7.

Remove the two screws shown below that secure the battery bracket, and remove the battery assembly from the enclo-
sure.

8.

Install the new battery assembly in the enclosure using the two screws removed in Step 7.  Plug in the battery connec-
tor.

Battery connector 

Battery and

bracket 

9.

Carefully position the PC board in the enclosure and loosely install the four screws on the bottom of the enclosure.
Loosely install the two screw locks on the RS-232 connector.  After all of the screws are started, tighten the four screws
on the bottom of the enclosure and the two screws on the RS-232 connector.

Screw

locks

10.

Check the ribbon cable connector and make sure the retainer clip is in the “up” position.

3

Summary of Contents for LI-3000C

Page 1: ... recycled or disposed of properly Follow these steps to replace the rechargeable battery pack in the LI 3000C 1 Remove the four screws on each side of the LI 3000C top panel keypad 2 Carefully lift the top panel away from the enclosure be cautious of the ribbon cable that is connected to the main PC board 3 Lift the retainer clip on the ribbon cable connector and pull the ribbon cable up and out o...

Page 2: ...ow from the bottom of the instrument enclosure 5 Remove the two screw locks from the RS 232 connector using a nutdriver or small wrench 6 Carefully lift the PC board out of the enclosure and unplug the battery connector Battery connector 2 ...

Page 3: ...ry connector Battery and bracket 9 Carefully position the PC board in the enclosure and loosely install the four screws on the bottom of the enclosure Loosely install the two screw locks on the RS 232 connector After all of the screws are started tighten the four screws on the bottom of the enclosure and the two screws on the RS 232 connector Screw locks 10 Check the ribbon cable connector and mak...

Page 4: ...op X line such as X Y Use the numeric keys to enter the current time Press to erase an incorrect character Press ENTER when fin ished e The display returns to the Set Clock function Press ENTER to confirm that the new date is displayed Press ENTER again to confirm that the new time is displayed Press the AREA key to exit the MENU functions Use the numeric keys to enter the current date Press to er...