LI-COR LAI-2000 Short Manual Download Page 2

2.

 

Set the operation mode. Press SETUP key, position 11 Set Op Mode on the top line, 
ENTER. Use 

 or 

 to display ‘1 sensor X’ on the top line, ENTER. You are then 

prompted for the sequence of above and below readings: 

 means above canopy 

reading, 

 means below canopy reading. If needed, set a new sequence and press 

ENTER. Choose the number of repetitions (reps) which is the number of times you 
whish to repeat the pattern for one complete measurement. 

 

3. Logging data and computing LAI.  

 
Press LOG, this initiates the process of collecting the necessary above and below 
measurements to calculate LAI. 
Respond to the two prompts (1) What (which crop) and (2) Where (plot) 
 
N.B.:  While logging in ‘1 sensor X’ mode, one line of the display is in real time, the other is 

the summary line. These lines will change positions with the position of the real time 
line indicating whether the next reading is to be an above canopy, or below canopy 
reading See figure 2. The * is always on the real time line. When the real-time line is 
on the top of the display, the number to the left of the * indicates the number of above 
canopy readings thus far. When the real-time line is on the bottom of the display, the 
number to the left of the * indicates the number of below canopy readings thus far. If 
wanted, the channel ID can be changed by 

 or 

 ( or by 

 or 

 if the real-time line 

is at the bottom of the display). This is only needed when more than one optical sensor 
is connected.  
The summary line displays three numbers summarizing the measurement. (1) the 
number of the above/ below pairs that have been accepted (SMP) (2) The running 
average of LAI (LAI) (3) the standard error (SEL) 
 

 

Figure 2: Display in logging modus 

 
 

Measurement 

 

1.

 

Hold the sensor leveled just above the canopy.  

2.

 

Press ENTER or button on the sensor (beep sounds) 

3.

 

if the measurement is finished, a 2

nd

 beep sounds.  

4.

 

When the last measurement has been logged, the display shows COMPUTING… 

 

View measurement results 

 
After logging, the viewing mode is automatically activated. If not, press FCT 27. 