LI-COR 5700-DS Accessories Manual Download Page 15

Chapter 2:

Pearl Clean Box

Description

Figure 2-1. 9300-022 Pearl Clean Box.

The purpose of the Pearl Imager Clean Box is to
provide a HEPA-filtered closed environment for
mice requiring added protection from exposure
to pathogens.

The lower portion of the Clean Box is an
imaging bed with a heater plate similar to the
standard Pearl imaging bed. The heater plate
can be used to maintain the body temperature
of the mouse when the Clean Box is locked in
place on either the Pearl Imager or Pearl
Docking Station. The handle of the Clean Box is
used to lock it in place using the same “rotate
down” method as the standard imaging bed.
When locked in place, anesthesia gas will be
supplied to the mouse via the standard Pearl
nose cone. HEPA filters are provided for both the Pearl Imager and Pearl Docking Station to protect
the mouse by filtering the anesthesia gas. When the Clean Box is undocked, the gas ports
automatically close.

The Pearl Clean Box window is a specially coated cell cast acrylic window for optimum light
transmission and durability. The coating on the exterior surface of the window will help minimize
smudging and scratching. The interior surface is coated to minimize scratches.

Important:

Do not discard the clear protective cover for the Clean Box during unpacking. The cover

protects the imaging window of the Clean Box during storage and transfer from the Docking
Station to the Pearl Imager.

Installation

Installation consists of installing HEPA filters and installing a nose cone. The Clean Box should
also be sanitized before initial use as described later in this chapter.

Connecting a HEPA Filter to the Docking Station

When using the Pearl Clean Box, one of the included HEPA filter assemblies should be installed
between the anesthesia supply tube and the docking station rotameter as described in

Assembling the

Pearl Docking Station

on page 2, Step #5).

Page 9

Summary of Contents for 5700-DS

Page 1: ...Pearl Trilogy Accessories Manual...

Page 2: ......

Page 3: ...This product model number 5700 DS is a CE marked product For conformity information contact LI COR Support at http www licor com biotechsupport Outside of the U S contact your local sales office or d...

Page 4: ...nt NOTE Notes contain additional information and useful tips IMPORTANT Information of importance to prevent procedural mistakes in the operation of the equipment or related software Failure to comply...

Page 5: ...d by copyright All rights are reserved No part of this document may be photocopied reproduced or translated to another language without prior written consent of LI COR Inc Printing History Publication...

Page 6: ...Box Description 9 Installation 9 Connecting a HEPA Filter to the Docking Station 9 Connecting a HEPA Filter to the Pearl Imager 10 Nose Cone Installation and Replacement 10 Sanitizing the Pearl Clean...

Page 7: ...on trol to a Pearl accessory imaging bed outside of the Pearl Imager allowing animals to be prepared for imaging at an external station l Warms the heater plate in any of the Pearl accessory imaging b...

Page 8: ...instructions dans ce chapitre pour vous assurer que le port d chappement est reli un filtre charbon que le syst me d anesth sie fourni par l utilisateur est correctement reli et que le cone de nez est...

Page 9: ...UST on the rotameter to the port labeled OUT on the back of the docking station 5 If using the Pearl Clean Box connect one of the provided HEPA filter assemblies packaged with the Clean Box to the por...

Page 10: ...rage de Pearl n est compatible qu avec les syst mes d anesth sie qui fournissent le gaz isoflurane Le d bit et la pression dans l instrument ne devraient pas exc der 1 litre par minute et 6 9 kPa resp...

Page 11: ...UST port Recommendations for monitoring and replacement of the charcoal filter are given in Chapter 3 of the Pearl Imager Operator s Manual Figure 1 6 The included charcoal filter or a user supplied c...

Page 12: ...structions for the specific imaging bed you are using WARNING Make sure a black nose cone plug is tightly inserted in the nose cone before turning on the anesthesia gas Failure to insert a nose cone p...

Page 13: ...the adjustment knob on the rotameter to set the flow rate to 0 5 liters per minute typical and start gas flow from isoflurane vaporizer or equivalent Allow 1 minute to prime the system with isoflurane...

Page 14: ...ire or loose connections Periodically inspect external tubing for cracking or damage that could cause leaks Docking Station Specifications Docking Station Operating Conditions For Indoor use only oper...

Page 15: ...the anesthesia gas When the Clean Box is undocked the gas ports automatically close The Pearl Clean Box window is a specially coated cell cast acrylic window for optimum light transmission and durabi...

Page 16: ...ion after it is plugged in 3 Connect the opposite end of the HEPA filter assembly to the provided anesthesia supply tube provided with the Pearl Imager Nose Cone Installation and Replacement Nose cone...

Page 17: ...the imaging bed surface with 0 5 chlorine bleach Ethanol or isopropanol can be also used as a cleaning solution for applications where chlorine bleach is not required If necessary clean any fingerprin...

Page 18: ...IMPORTANT Move the Clean Box quickly Anesthesia gas does not flow while the Clean Box is undocked 8 Remove the Clean Box cover to expose the window close the imaging drawer and acquire an image as des...

Page 19: ...rove insufficient use LI COR Optical Tissue and LI COR Optical Cleaner Apply a few drops to an optical tissue and gently clean the window with a circular motion Follow with a second dry tissue to remo...

Page 20: ...iosciences GmbH Siemensstra e 25A 61352 Bad Homburg Germany Phone 49 0 6172 17 17 771 Fax 49 0 6172 17 17 799 bio eu licor com LI COR Biosciences UK Ltd St John s Innovation Centre Cowley Road Cambrid...

Reviews: