Trigger Cable
•
Trigger Cable is connected between control box and X-ray Generator, and used only for manual mode, not auto mode.
AC-IN T4L 250V
Ethernet Sync
DXD 1 DXD 2
Labeling of pins
X-ray Generator
NC: No Connection
No.
Labeling of pins color
Description
1
Red
Power: X-ray Generator Supply Voltage (DC 12V ~24V)
Use
2
Brown
Expose signal from Generator to Control Box
Use
3
Blue
Cancel REQ Ground
NC
4
Orange
Prepare Acknowledge signal from Control Box to Generator
Use
5
Yellow
Expose Acknowledge signal from Control Box to Generator
Use
6
Green
Prepare signal from Generator to Control Box
Use
7
Black
Cancel request VCC
NC
8
Gray
Prepare Acknowledge Ground
NC
9
Violet
Ground of Signals
Use
NOTE
•
Connection of trigger cable and X-ray generator is to be performed by qualified personnel. The description of each pin is in the common language for this industry.