ADVANCED MICRO CALIBRATION BATH TEMPERATURE
LCA 30/50
www.leyroinstruments.com
25
11.2.1. Cleaning of calibrators with inner cap
In gauges with insert bushing a small amount of metal powder is produced, which can clog the
nozzle and the bushing. To prevent this, disassemble the inner bushes periodically and before any
long period out of service. C lean the perforation of the micro bath with pressurized air and with a
dry cloth the perforation and the cap.
11.2.2. Cleaning of the fan grille
All the micro-bathrooms have a fine mesh grid inside, through which cooling air enters the micro-
bath. Clean the grille regularly depending on air pollution with a vacuum cleaner or brush
11.2.3. Internal cleaning of the Calibration bath
Remove all the silicone from the tank. Remove the sensor basket from the tank and clean the tank,
the magnetic stirrer and the tank by applying water with a large amount of cleaning substances.
Allow all components to dry. If distilled water is used, remove the calibration fluid and let dry the
sensor basket, the magnetic stirrer and the tank
11.2.4. External cleaning of the Calibration bath
Clean the outside of the calibration micro-bath with a damp cloth and a little water or with a non-
aggressive cleaning product without solvent.
11.3. Re calibration
ISO /17025 Certificates
The calibration micro-bath has been adjusted and checked before shipment using standard quality
internationally recognized measuring instruments. According to ISO / 17025 the calibration micro-
bath has to be checked at appropriate periodic intervals depending on the use. It is recommended
to calibrate the instrument by the manufacturer at periodic intervals of approximately 12 months or
every 500 hours of approximate operation. All Factory Re-calibration also includes an exhaustive
and free check of all system parameters in terms of specifications. Any deviation from the basic
values is corrected. The bases of the Re-calibration are the guidelines of ISO / 17025. The
measurements detailed in this document must be observed and applied during the Re calibration.
12. Accessories included LCB 30/50
Syringe for the insertion or extraction of fluid
1,5 mm network cable with connector type F according to CEE7/4
Basket
Magnetic stirrer and metal screw cap.
12.1. Additional options
Embossed aluminium case with handle and reinforced wheels.
Methacrylate cap with perforations
13. PASSWORD
The access code to the equipment to configure the different options is:
leyro30
Summary of Contents for LCA 30
Page 6: ...ADVANCED MICRO CALIBRATION BATH TEMPERATURE LCA 30 50 www leyroinstruments com 4 ...
Page 30: ...ADVANCED MICRO CALIBRATION BATH TEMPERATURE LCA 30 50 www leyroinstruments com 28 ...
Page 32: ...ADVANCED MICRO CALIBRATION BATH TEMPERATURE LCA 30 50 www leyroinstruments com 30 ...
Page 34: ...ADVANCED MICRO CALIBRATION BATH TEMPERATURE LCA 30 50 www leyroinstruments com 32 ...
Page 36: ...ADVANCED MICRO CALIBRATION BATH TEMPERATURE LCA 30 50 www leyroinstruments com 34 ...