Connection
13
2 Connection
The unit may only be connected by a qualified electrici-
an and in accordance with the local regulations.
Warning
Inside the TURBOTRONIK there are hazar-
dous voltages. Open the TURBOTRONIK
only after disconnecting the mains.
2.1 Selecting the mains vol-
tage (optional)
The mains voltage setting of the TURBOTRONIK can be
changed; see „Technical data“.
Caution
The TURBOTRONIK is damaged when
connected to the mains with an incorrect
mains voltage setting.
Open the hinged cover (3/3). Remove the printed circuit
board (3/5) and plug it back in such that the appropriate
mains voltage value is visible.
If the voltage drops below 216 V in 230 V mains select
the 220 V setting.
If necessary change the fuse.
Required fuse:
100-120 V
10 A
220-240 V
5 A
Close the hinged cover again. The set mains voltage is
visible from the outside.
TURBOVAC
100
220
240
120
100/120V F10A
220/240V F5A
START STOP RELAIS
1 2 3 4 5 6
HV
PE
1
2
3
4
5
6
7
Key to fig. 4
1 Plug for the TURBOVAC
connection line
2 Earth connection
3 Mains voltage settings and
fuse; see fig. 3
4 Terminals for remote
control; see fig. 5
5 Terminals for relays
6 Fuse
7 Plug for the mains cable
Fig. 4 Rear of the unit