LEYBOLD PS 113 A Operating Manual Download Page 6

Safety, Technical Data 

 
 

6

 

GA09616_002_C0 – (2016-10) – 

 Leybold 

 

 

1.4 

Liability and Warranty 

Leybold assumes no liability and the warranty becomes null and void if the end-
user or third parties 

  disregard the information in this document 

  use the product in a non-conforming manner 

  make any kind of changes (modifications, alterations etc.) to the product 

  use the product with accessories not listed in the corresponding product 

documentation. 

 
 
 
 
 
 

 

2 Technical 

Data 

 

Switching pressure 

6 mbar 
(below atmospheric pressure) 

Return switching pressure 

3 mbar 
(below atmospheric pressure) 

Switching inaccuracy 

±2 mbar 

Max. operating pressure (abs.) 

2000 mbar 

Operating temperature 

0 to 85 °C 

Helium permeation 

<1·10

-6

 mbar·l·s

-1

 

Switching contact 

changeover contact 

Switching capacity 

100 mA / 24 VAC 
30 mA / 24 VDC 

Electrical connection 

standard cable 3 m 

Vacuum connection 

DN 16 KF 

Protection type 

IP 44 

Materials in contact with the medium stainless 

steel 

1.4305, 

1.4310 

stainless steel 1.4300 PTFE coated 
EPDM 

Internal volume 

2 cm

3

 

Weight 

315 g 

 

 

Dimensions [mm] 

 

Fig. 1  Dimensional diagram

 

PS 113 A 

 

All manuals and user guides at all-guides.com

all-guides.com

Summary of Contents for PS 113 A

Page 1: ...PS Low Operat Part Num 230 011 S 11 w Pres ing Man ber 13 A ssure ual GA0 A Safet 09616_00 ty Sw 02_C0 witch All manuals and user guides at all guides com a l l g u i d e s c o m...

Page 2: ...m systems It is used e g to automatically interrupt the gas supply when venting vacuum systems with a purge gas at a pressure of 3 mbar below atmos pheric pressure Unpacking and Checking Unpack the lo...

Page 3: ...ric pressure an elastic diaphragm actuates a changeover contact which in turn may be used to directly switch any ancillary equipment The electrical connections are placed under a plastic cover Scope o...

Page 4: ...ns 5 1 3 General Safety Instructions 5 1 4 Liability and Warranty 6 2 Technical Data 6 3 Installation 7 3 1 Vacuum Connection 7 3 2 Electrical Connection 7 3 3 Start up 7 4 Maintenance 8 5 Returning t...

Page 5: ...y experience or who have been instructed by the end user of the product 1 3 General Safety Instructions Adhere to the applicable regulations and take the necessary precautions for the process media us...

Page 6: ...6 mbar below atmospheric pressure Return switching pressure 3 mbar below atmospheric pressure Switching inaccuracy 2 mbar Max operating pressure abs 2000 mbar Operating temperature 0 to 85 C Helium p...

Page 7: ...PS 113 A is supplied with a connected 3 m long standard cable The length of this cable may be changed The electrical connection terminals and the strain relief are accessible after un screwing the cr...

Page 8: ...substances e g radioactive toxic caustic or microbiological Adhere to the forwarding regulations of all involved countries and forwarding companies and enclose a completed declaration of con taminati...

Page 9: ...of in accordance with special regulations Dispose of such products in accordance with the relevant local regu lations Separating the components After disassembling the product separate its components...

Page 10: ...Declaration of Contamination 10 GA09616_002_C0 2016 10 Leybold Declaration of Contamination All manuals and user guides at all guides com...

Page 11: ...EU Declaration of Conformity GA09616_002_C0 2016 10 Leybold 11 EU Declaration of Conformity All manuals and user guides at all guides com a l l g u i d e s c o m...

Page 12: ...0 Les Ulis cedex Sales and Service T 33 1 69 82 48 00 F 33 1 69 07 57 38 info ctb leybold com sales ctb leybold com Leybold France S A S Valence Factory 640 Rue A Berg s B P 107 F 26501 Bourg l s Vale...