Maintenance and Service 

 

Danger of possibly contaminated parts! 
Contaminated parts can cause personal injuries. Inform your-
self regarding possible contamination before you start working. 
Be sure to follow the relevant instructions and take care of 
necessary protective measures. 

 
The unit requires no maintenance. External dirt and soiling can be removed by a 
damp cloth. 
 
Should a defect or damage occur on your DI2000, please send the instrument to 
us for repair and enclose a decontamination declaration. 
 
 
 

 

The unit is not prepared for customer repair! 

 
 
 
 
 
 

 

 
Malfunction of the unit, which is caused by contamination, wear 
and tear or a broken sensor is not covered by warranty. 
 

 
 
 
Error messages and malfunction 
 

 

Problem 

Possible Cause 

Correction 

high measurement error  

contamination, ageing, 
extreme temperature, 
maladjustment 
 

readjustment 

output signal <= 3,8 mA (DI)  
output signal <= 1,5 V (DU) 
 

defective sensor 
 

return sensor for repair 
 

output signal >= 21 mA (DI) 
output signal >= 10,5 V (DU) 
 

pressure overrange 
or defective sensor 
 

actual pressure too high 
or return sensor for repair  

adjustment not possible 

measurement error exceeds 
possible range of 
readjustment 
 

return sensor for repair 
 

 
 

Summary of Contents for 15813V01

Page 1: ...Vacuum Transducer DI2000 DI2001 DU2000 DU2001 Operating Instructions 300360490_002_C0 Cat No 15813V01 DI2000 Cat No 15815V01 DI2001 Cat No 230502V01 DU2000 Cat No 230503V01 DU2001...

Page 2: ...ctrical Connection 7 3 3 1 Operation with Display Unit 7 3 3 2 Operation with Other Supply And Evaluation Units 8 4 Operation 9 4 1 General 9 4 2 Adjustment 10 4 2 1 Zero Adjustment 10 4 2 2 Zero Adju...

Page 3: ...ed Consider possible reactions between materials and process media Consider possible reactions of the process media due to the heat generated by the product Do not carry out any unauthorized conversio...

Page 4: ...put signal Unless these different features are described explicitly only model DI2000 is exemplarily mentioned in the following 2 2 Delivery Content Included in the delivery consignment are Transducer...

Page 5: ...ers use EPDM as sealing material The transducer can be mounted to suitable flange connectors or by means of the G1 4 female thread Proper Use The DI2000 serves exclusively to provide total pressure me...

Page 6: ...cleanliness and damage prevention when using vacuum compo nents Remove the protective cover is required again during maintenance work Make vacuum connection via small flange DN16 ISO KF If you use a...

Page 7: ...d of small flange connec tions 2 5 bar abs Use clamps that are suitable for overpres sure operation 3 3 Electrical Connection 3 3 1 Operation with Display Unit The transducer can be connected to suita...

Page 8: ...wn below DI2000 DI2001 Pin description connector topview Diode plug 7 pin male Pin 1 Pin 2 brown white brown Pin 3 blue white blue Pin 4 Pin 5 green white green Pin 6 orange white orange Pin 7 Schirm...

Page 9: ...easure for the applied pressure Output Signal The analog output signal of the transducer is available immediately after the voltage supply has been connected The output of 4 0 20 mA of a DI2000 DI2001...

Page 10: ...tment The adjustment at zero pressure is an offset compensation of the measurement signal For zero adjustment real pressure must be less than 0 1 mbar An adjustment at zero pressure is only possible i...

Page 11: ...d be 0 mbar For adjustment first remove the rubber caps from both buttons 1 If the pressure reading is 0 in case of a DISPLAY unit or ccc0 cc0 or c0 in case of a CENTER unit the output signal is too l...

Page 12: ...or an interval of 5s the adjusted value is set Now insert the rubber cap again 4 2 4 Full Scale Adjustment DU2000 DU2001 With Display Unit The full scale adjustment is a span correction of the measure...

Page 13: ...and enclose a decontamination declaration The unit is not prepared for customer repair Malfunction of the unit which is caused by contamination wear and tear or a broken sensor is not covered by warra...

Page 14: ...3 ceramic FPM DI2000 DU2000 EPDM DI2001 DU2001 Operating temperature 0 60 o C Storage temperature 40 70 o C Response time 20 ms Voltage supply DI 12 30 VDC two wire DU 14 5 30 VDC four wire Electrical...

Page 15: ...300360490_002_C0 09 2016 Leybold 15 Declaration of Conformity...

Page 16: ...Headquarter Leybold GmbH Bonner Strasse 498 D 50968 Cologne T 49 0 221 347 0 F 49 0 221 347 1250 info leybold com w w w l e y b o l d c o m...

Reviews: