LEYBOLD 157 89 Manual Download Page 14

Page 4
Correction for the type of gas
GAS.n2 

Air / Nitrogen (N

2

)

GAS.Ar 

Argon

Setting when shipped:

Air / Nitrogen

For a more accurate determination of the true pressure
for gases other than air or N

2

in the system, the display-

ed value has to be multiplied by a factor which is cha-
racteristic for the other type of gas.

Gas

Correction factorr

————————————————————————
O

2

0,88

The correction factors 

He

4,7

refer to the setting 

H

2

2,28

for GAS.n2 in each case.

Ne

2,16

Page 5
Software release number and locking (LOCK); refer also
to Section 2.3.9.2.

The transition from the parameter mode to the normal
measurement mode is made by pressing the key of the
desired measurement channel (TM 1, TM 2 or PM) or by
pressing the PARA key.

Parameter level 2
Parameter level 2 
is called up by pressing the incre-
ment 
(2/13) key when parameter page 5 is being dis-
played.

Parameter level 2 contains the following:

Page 6
Units of measurement TORR, PA, MICRON, MBAR
The currently active unit flashes.

Note
The unit which is set up here, applies to all three connec-
ted measurement channels.

Supplied condition:

MBAR (230 V model)
TORR (100 V or 120 V model)

Even when selecting the unit „MICRON“ for a TM chan-
nel, the pressure readings for the PM channel will always
be in „TORR“.

If the TM channels are to be set to „MICRON“ this can
not be done via the PM channel. This is only possible
when having previously selected a TM channel.

Page 7

Note
On parameter page 7 of instrument parameter level 2
there are different displays for the THERMOVAC channel
and the PENNINGVAC channel.

Page 7 for the THERMOVAC channel
Filament material of the connected sensor (FILAMENT).
FIL tu Tungsten TR 211, TR 201 and TR 205
FIL ni Nickel TR 206
FIL pt Platinum TR 216

Setting when shipped: FIL tu

Page 7 for the PENNINGVAC channel
Instrument parameter page 7 is used for automatic, pres-
sure dependent switching of the PENNINGVAC channel.

After calling up instrument parameter page 7 the display
will indicate AUTO.0 or AUTO.1

AUTO.0

The high voltage may be switched on and off
via the „HV“ key or via the external input HV-
ON. The status of the PENNINGVAC channel
is not controlled via the measurement chan-
nel TM 2.

AUTO.1

Depending on the pressure the high voltage
is switched on and off through measurement
channel TM 2. The high voltage may be swit-
ched off and on at any time by pressing the
„HV“ key or through the external input HV-
ON, provided the pressure is below 1·10

-2

mbar.

The threshold pressure for
- switching on the high voltage is:

< 1·10

-2 

mbar = 7.5·10

-3

Torr = 1 Pa = 8 Micron

- switching off the high voltage is:

> 5·10

-2

mbar = 3.75·10

-2

Torr = 5 Pa = 37 Micron

In the case of automatic operation, the pressure display
is switched to the best measurement channel (TM 2/PM)
depending on the pressure. In the event of a faulty sen-
sor in channel TM 2 the high voltage has to be switched
on by the operator.

The display switches from the THERMOVAC channel to
the PENNINGVAC channel at a pressure of < 3·10

-3

mbar (2.5·10

-3

Torr) and from the PENNINGVAC channel

to the THERMOVAC channel at a pressure of > 5·10

-3

mbar.

Having actuated the keybord the change-over can be
delayed by 1 min.

Setting when shipped:

AUTO.1

Page 8
Equipment parameter page 8 is used for switching the
numerical display from standard resolution (2-digit man-
tissa) to high resolution (3-digit mantissa).

Std

Standard resolution

3.0  

-2

HIGH

High resolution

3.00

-2

For this refer to tables 3 and 4 in Section 2.10.1 or tables
7 and 8 in Section 2.10.2.

14

Summary of Contents for 157 89

Page 1: ...COMBIVAC CM 31 Cat No 157 89 896 89 897 89 Operating Instructions Vacuum Pumps Instrumentation Fittings and Valves LEYBOLD VACUUM GA 09 504 5 02...

Page 2: ...Electrical connection 8 2 2 1 Changing the line voltage setting and exchanging the fuses 9 2 3 Controls and their functions 9 2 3 1 Bar graph display 9 2 3 2 Digital display 10 2 3 3 Measurement unit...

Page 3: ...format for printer output 27 3 3 2 3 Output rate for printer output 27 3 3 2 4 End character 27 3 4 Data output and data formats 27 3 4 1 Measurement data output 27 34 1 1 Remote control operation 27...

Page 4: ...e The COMBIVAC CM 31 is a universal vacuum gauge which combines two principles of measurement Pirani Thermovac and cold cathode Penningvac for the measurement and control of vacuum pressures within th...

Page 5: ...indicating cir cuit with n o contact Max switching voltage 250 V AC 60 V DC Max switching capacity 5 A AC resistive load 0 7 A DC Contact life 60 000 cycles Setting range of the variable thresholds P...

Page 6: ...ntrol arrange ments The two THERMOVAC channels are immediately active as soon as the line voltage is applied The PEN NINGVAC channel may be switched on and off through the second THERMOVAC channel TM...

Page 7: ...erminal strips 200 60 806 One 5 way screw terminal strip 200 60 807 Three 8 way screw terminal strips 200 60 808 Four screws M 3 x 8 mm 200 80 029 Support stand for table use 200 60 900 Four adhesive...

Page 8: ...t or after having exchanged a sensor TM channel TEST and noSEn will be displayed alternatingly for 1 to 5 seconds While this is in progress the instrument is trying to determine which THERMO VAC senso...

Page 9: ...ntrols and the dis play elements is given in Fig 2 The instrument is operated via 7 keys Note When pressing a key which has no function in that parti cular operating mode symbol 2 9 comes on 2 3 1 Bar...

Page 10: ...ressure display Trigger and equipment modes are indicated in the status display area 3 1 These are related to the keys below which are used to select the measurement channel An overview giving the arr...

Page 11: ...not be active However if these conditions have been met once after switching on the instrument will indicate that it is ready Thereafter the triggers are active independently of the pressure cable or...

Page 12: ...PARA 5 1 key is pressed when wanting to check or change the settings of the various instrument parame ters The PARA message in the status display area comes on and the first parameter page of the cur...

Page 13: ...modes L Level trigger Both trigger outputs are operated independently of each other Thresholds may be set up either within the range between 1 10 8 and 1 10 2 mbar for PM measurement channel or for T...

Page 14: ...sten TR 211 TR 201 and TR 205 FIL ni Nickel TR 206 FIL pt Platinum TR 216 Setting when shipped FIL tu Page 7 for the PENNINGVAC channel Instrument parameter page 7 is used for automatic pres sure depe...

Page 15: ...characteristic 0 0 1 1 5 V 10 V Anl 7 1 10 9 1 10 8 1 10 7 mbar Anl 6 1 10 8 1 10 7 1 10 6 mbar Anl 5 1 10 7 1 10 6 1 10 5 mbar Anl 4 1 10 6 1 10 5 1 10 4 mbar Anl 3 1 10 5 1 10 4 1 10 3 mbar Anl 2 1...

Page 16: ...he gauge head connector Those terminal strips which are above each other are always related to one THERMOVAC channel The wiring is the same for both measurement channels but the numbering of the pins...

Page 17: ...0 for external contact on 211 211 HV control input for PM HV ON 212 Ground for HV control input 231 Chart recorder output 0 to 10 V REC 232 Ground for chart recorder output For the PENNINGVAC channel...

Page 18: ...The required panel cut out is given in Fig 8 2 5 4 Using the CM 31 as a table top instrument When using the CM 31 as a table top instrument a sup port stand Ref No 200 60 900 may be fitted to the bot...

Page 19: ...ied out as follows Vent the vacuum system and adjust the 100 potentio meter on the THERMOVAC gauge head so that the fol lowing bargraph display is obtained Note In order to ensure a stable but none th...

Page 20: ...s or a brief mains failure 1 to 3 seconds for example In this case the instrument or the affected mea surement channel is reset to a stable through a Watchdog function Display Status FAULT all other s...

Page 21: ...open Chart recor der output 10 2 10 6 V 0 to 10 V 10 2 10 6 V 10 2 10 6 V mbar Pa Torr Micron Chart rec output volt 1 0 10 3 1 0 10 1 1 0 10 3 1 0 00 V 2 0 10 3 2 0 10 1 2 0 10 3 2 0 50 V 5 0 10 3 5...

Page 22: ...r mode and standard resolution mbar Pa Torr Micron Measurement Trigger hysteresis uncertainty Level trigger 1 0 10 3 to 4 9 10 3 1 0 10 1 to 4 9 10 1 1 0 10 3 to 4 9 10 3 1 2 3 4 20 nicht einstellbar...

Page 23: ...put 10 2 10 6 V 10 2 10 6 V 0 2 10 2 V 10 2 10 6 V mbar Pa Torr Chart rec output volt 1 0 10 9 1 0 10 7 1 0 10 9 0 00 V 2 0 10 9 2 0 10 7 2 0 10 9 0 43 V 5 0 10 9 5 0 10 7 5 0 10 9 1 00 V 9 0 10 9 9 0...

Page 24: ...0 05 increments 1 00 10 3 to in 0 1 increments in 0 1 increments 6 00 10 3 to in 0 2 increments in 0 2 increments 1 00 10 2 Table 7 Display resolution and display increments Table 8 Assignment Measur...

Page 25: ...ire an input buffer size of at least 30 characters XON and XOFF handshaking is not used to control the data flow and will lead to error messages The RS 232 C interface requires at least 3 lines Transm...

Page 26: ...erface will usually not operate properly 3 3 1 4 Acknowledgement character In the remote control mode the A Series instrument will respond to each string it receives and which has been terminated by t...

Page 27: ...ter In the printer mode the character string CR LF carriage return and line feed is used as the end charac ter see Section 3 3 1 3 3 4 Data output and data formats 3 4 1 Measurement data output 2 4 1...

Page 28: ...kely to occur When the A series instrument is just processing a com mand or transmitting a string while further characters are being transmitted to the instrument it will ignore these as invalid 3 5 I...

Page 29: ...nsor see Tech nical Data Thresholds Specifically n nn Mantissa possibly with additional sign mm Exponent always with sign Example TRC W TM1 1 00 2 00 3 5 3 Operating parameters Entry of parameters via...

Page 30: ...thout test LINE INPUT 1 FailMessg get error code from CM31 LOCATE 4 1 PRINT SPACE 79 clear screen line LOCATE 4 1 PRINT RS232 failure on cold cathode high voltage on command LOCATE 4 40 PRINT FailMess...

Page 31: ...ype argon LOCATE 9 40 PRINT FailMessg error code to screen ELSE LOCATE 9 1 PRINT SPACE 79 clear screen line LOCATE 9 1 PRINT setting gas type argon successful END IF read current gas type of CM31 PM c...

Page 32: ...NE INPUT 1 AckNakTest get handshake character without test LINE INPUT 1 FailMessg get error code from CM31 LOCATE 15 1 PRINT SPACE 79 clear screen line LOCATE 15 1 PRINT failure on setting of TM1 no 1...

Page 33: ...r value to screen END IF take measurement value from CM31 channel TM1 PRINT 1 MESr Tm1 command read measurement value from CM31 TM1 LINE INPUT 1 AckNakTest get handshake character from CM31 IF AckNakT...

Page 34: ...PC IBM PC Remark for A side Desig Pin Pin Desig Remark for PC side leave unconnected 1 8 DCD possibly connect to 6 link TxD 2 3 RxD link link RxD 3 2 TXD link 4 20 DTR link GND 5 7 GND link link DTR 6...

Page 35: ...form which we shall send to you upon re quest A copy of this form is printed at the end of the Operating Instructions Declaration of Contamination of Vacuum Equipment and Components Either fasten this...

Page 36: ...5 Brief operating instructions 36...

Page 37: ...easurement channel ADisplay resolution STD HIGH For TM measurement channel Cable length alignment for TM CLA CL xxx For PM measurement channel Setting of the analogue output log lin log Ano 8 lin Anl...

Page 38: ...38...

Page 39: ...gue numbers 157 89 896 89 897 89 157 90 The product meets the requirements of the following directives EEC Directive on Low Voltages 73 23 EWG LHK GI 0008 02 02 95 LEYBOLD AG Vacuum Technology Bonner...

Page 40: ...1 80 5 676 30 RSP 09 95 Printed in Germany on chlorine free bleached paper LEYBOLD AG Vakuum Bonner Stra e 498 Bayenthal D 50968 K ln Telefon 0221 347 0 Telefax 0221 347 1250...

Reviews: