User Manual for PuraLev
®
i30SU
www.levitronix.com
PL-4051-00, Rev03, DCO# 20-275
30
4.4 Mechanical Installation
– General Instructions
!
WARNING
The driver is not rated to be placed next to flammable gases. Do not
use the driver next to flammable gases without adequate safety
precautions to fulfill the relevant regulatory requirements.
4.5 Mechanical Installation of Stand-Alone and EasyConnect Model
The
Stand-Alone
and
EasyConnect
model are delivered with the mounting base plate
MBP-i30.1
, which can
be used to fix the driver with four screws on the mounting slots of the base plate (see
EasyConnect
model). Mounting of the driver can be done in either the
horizontal or vertical position.
When mounting the pump head fittings into a circuit it has to be taken care that no mechanical stress is
acting on the fittings, which can result in distorting creeping effects.
The outlet orientation of the pump head can be changed in steps of 45° by assembling the pump head
socket accordingly (see
).
Figure 28: Vertical and 45° orientation of pump head with rotation of pump head socket
4.6 Mechanical Installation of OEM Model
If the mounting base plate
MBP-i30.1
is used, the driver can be fixed with four screws on the mounting slots
). Mounting of the driver can be done in either the horizontal or vertical
position.
For other mounting configurations the 2 screws on the side of the driver are an option to be used (see
When mounting the pump head fittings into a circuit it has to be taken care that no mechanical stress is
acting on the fittings, which can result in distorting creeping effects.
The outlet orientation of the pump head can be changed in steps of 45° by assembling the pump head
socket accordingly (see
).
Vertical Pump Head Outlet
45° Pump Head Outlet
Pump Head Socket