User Manual for PuraLev
®
200SU
www.levitronix.com
PL-4030-00, Rev06, DCO# 20-144
7
2.2.3 ATEX / IECEx System Configuration
Together with the standard pump-head and controllers an
ATEX / IECEx
certified motor allows installation of
motor and pump head within an
ATEX Zone 2
). The
ATEX / IECEx
motor (
Pos. 2b
in
) comes delivered with special connectors and according extension cables (see
). One option
to lead the cables outside of the
Ex
area is an
Ex
certified cable sealing system as listed in
Figure 5: System configuration for ATEX IECEx applications
2.3 General Environmental Conditions
Usage
Indoor
1
Altitude
Up to 2000
m
Operating ambient temperature
0 to 40
0
C
Storage ambient temperature
(Extremes for Transportation)
-20 to 80
0
C
-20 to 70
0
C
for single-use pump head
Operating humidity range (relative humidity)
15
– 95
%
(non-condensing)
Storage humidity range (relative humidity)
(Extremes for Transportation)
15
– 95
%
(non-condensing)
Normal storage conditions
Ambient temp.: 20 to 30
0
C
Relative humidity: 50
%
(non-condensing)
DC supply fluctuations
10% of nominal voltage
Transient over-voltages typically present on
the mains supply
Surge immunity according to EN 61000-4-5
(tested together with certified AC/DC power supply)
Pollution degree
2
Table 5: Environmental conditions for pump system
Note 1: Contact Levitronix
®
technical department (see Section 8) for outdoor usage.
Disposable
Pump Head
Fluid In
Fluid Out
Motor Power Cable
Adaptor/Extension Cable
Power Supply
24-48 VDC / 200 W
All
Controller
Types
ATEX conform cable sealing system
Cabinet Boundary for
ATEX Zone 2 Area
ATEX Zone 2 Classified Location
Chemical Resistance and IP-67 Space
ATEX Motor
Bayonet Socket
with Locking Pin
1
5
4
2b
8
3