![Levitronix PLD-iF30SU.1 User Manual Download Page 34](http://html.mh-extra.com/html/levitronix/pld-if30su-1/pld-if30su-1_user-manual_1886605034.webp)
User Manual for PuraLev
®
iF30SU
www.levitronix.com
PL-4078-00, Rev00, DCO# 20-228
34
5.2.2 Operation with Fieldbus (RS485)
The flow control driver can be operated with the fieldbus interface, which is an
RS485
with Modbus RTU
protocol. Contact
Levitronix
®
Technical Department
(see
Section 8
) for detailed protocol specifications. The
RS485 interface can be used in connection with the USB interface of a PC and the
Levitronix
®
Service
Software
. Various accessory cables are available to connect to the
RS485
of the drivers as illustrated in
EasyConnect
OEM
model.
5.2.3 Operation with User Panel LUI-B.1
The RS485 interface can also be used with the user panel
LUI-B.1
, which has the same menu functions as
the
Stand-Alone
model (described in
Section 5.1
). The relevant configurations are illustrated in
the
OEM
and in
EasyConnect
driver models.
5.3 Zero Adjustment of Flow Sensor
5.3.1 Stand-Alone Model
Zero adjustment of the flow sensor can be done manually in the system menu functions as described in
Section
. It is important to prepare the flow sensor as described in
Section 4.8
and to assure that the hydraulic
circuit is completely filled and zero flow is assured.
5.3.2 EasyConnect and OEM Model
Zero adjustment of the flow sensor can be started with the relevant digital input of the PLC (see
OEM
EasyConnect
model) or the
RS485
interface of the drivers (see
Section 5.2.2
With the user panel
LUI-B.1
zeroing can be done manually as for the
Stand-Alone
model (see
Section 5.1.2
It is important to prepare the flow sensor as described in
Section 4.8
and to assure that the hydraulic circuit is
completely filled and zero flow is assured.