User Manual for LEVIFLOW
®
LFSC-D Flowmeters
www.levitronix.com
PL-4507-00, Rev08, DCO# 20-024
24
4.2.2 Operation with PLC Interface
Table 9
shows the standard configuration of the electrical interface (PLC). For other configurations the
Levitronix
®
Service Software
shall be used (see reference on title page of this user manual).
is an example how to connect the
LEVIFLOW
®
converter to a relay to control a shutoff valve by a
digital output. Note that if protective diode is not in the relay, it has to be connected additionally as shown in
. This setup also shows how to connect a push-button to digital input (e.g. for reset of volume
counter, has to be configured by
Levitronix
®
Service Software
shows an example how to connect
digital in-/outputs and analog output of
LFC-1C-PC
to a
Levitronix
®
pump system.
Figure 18: Controlling a shutoff valve by digital output of LFC-1C-PC converter
Figure 19: Usage of PLC of LFC-1C-PC converter with a Levitronix
®
pump system
Shutoff Valve
Supply GND
Protective Diode
e.g. Diode 1N4007
24V
Push-Button
e.g. Relay Schrack
RY211024
1
DC24V+
2
DC24V-
6
DigOut 1+
7
DigOut 2+
8
DigOut -
9
DigIn +
10
DigIn -
5.1k
DC24V+
DC24V-
Leviflow LFC-1C
Converter LFC-1C
Levitronix Pump System
20
19
2
1
10
9
Supply GND
24V
Converter
LFC-1C-PC
Converter
LFC-1C-PC
Levitronix
Pump System