2.2 System Overview and General Specification
2.2.1 Stand-Alone System Configuration
The stand-alone pump system configuration
, consists of a controller with an integrated user panel
to set the speed manually.
The speed is automatically stored in the internal EEPROM of the controller. As an
option, the speed can also be set with an analogue signal.
Figure 3: Standard system configuration for standalone operation
(Speed setting with integrated user panel)
2.2.2 Extended System Configuration
The extended version of the pump system (
) consists of a controller with an extended PLC interface.
This allows setting the speed by an external signal and enables precise closed-loop flow or pressure control
in connection with either a flow or a pressure sensor. A USB interface allows communication with a PC in
connection with the
Levitronix
®
Service Software
. Hence parameterization, firmware updates and failure
analysis are possible.
Figure 4: Standard system configuration for extended operation
(Pressure or flow control with external sensor)
Chemical Resistance and IP67 Space
Disposable
Pump Head
Motor
Fluid In
Fluid Out
Motor Sensor Cable
Motor Power Cable
Adaptor/Extension
Cable for Sensors
Adaptor/Extension
Cable for Power
Power Supply
48 VDC / 600 W
Controller
for
Standalone
Operation
Stand Alone
Operation
Speed Control
PLC Interface
Speed Setting
- 1x Digital Input (Enable)
- 1x Digital Output (Status)
- 1x Analog Input (Speed)
Pumphead Socket
(Bayonet Type with Locking Pin)
Chemical Resistance and IP67 Space
Motor
Fluid In
Fluid Out
Motor Sensor Cable
Motor Power Cable
USB
- Service
- Firmware Update
- Configuration etc.
Flow- or
Pressure
Sensor
Adaptor/Extension
Cable for Sensors
Adaptor/Extension
Cable for Power
Extended PLC Interface
Speed, Flow
or Pressure Setting
- 4 Digital Inputs
- 4 Digital Outputs
- 4 Analoge Inputs
- 2 Analoge Outputs
Power Supply
48 VDC / 600 W
Controller
for
Extended
Operation
Disposable
Pump Head
Pumphead Socket
(Bayonet Type with Locking Pin)
Motor option with ETFE
coating and FEP cabling
Motor option with
Direct cabling
to controller
1
5b
4a
2a
2b
1
5b
4b
2b
2a
5a