LST-T004 Operations
05
LST-T004
LESITE
EN
5、Use screwdriver to tighten the 4 test sample compression screws (12)on the movable
plate() until the base material sample is completely clamped.
6、Check if the fixed slide bar(10), the drive shaft(11), and the movable plate (14) are free
of any foreign objects that hinder the operation; press the reset (4)button to reset the
previous residual data to zero; press the forward (6)button again,the machine starts to
run. You can see that 4 lines of data change on the display screen. When the base
material sample is completely broken, press the stop(8) button.
7、After the test, you can see the test data displayed on the display screen. Please confirm
whether the test sample is qualified according to the corresponding standard.
8、Loosen the 4 test sample compression screws (12)and take out the base material
sample ; if there is no test, please press the backward (7) button to return the movable
plate (14).
9、After the test is completed, turn off the power switch(1) and unplug the power plug. If
possible, please add some oil to the fixed slide bar (10) and the drive shaft(11).