5-2
5.2 Alarm Categories
The monitor provides two different types of alarms: physiological
alarms and technical alarms.
Physiological alarms
: also called patient status alarms, are
triggered by parameter measurement exceeding the set alarm
limits, or by an abnormal patient condition.
Technical alarms
: also called system status alarms, are
triggered by a device malfunction or the monitoring result
distortion due to improper operation or mechanical problems.
Apart from the physiological and technical alarms, the monitor can
also display messages describing the system status or patient status
in the prompt message area at the top of the screen.
For alarm events and prompt messages, see
5.3 Alarm Priorities
By severity, the alarms are classified into the following priority levels:
High priority alarms: indicates a life threatening situation or a
severe device malfunction. High priority alarms require an
immediate operator response.
Medium priority alarms: indicates abnormal vital signs or a
device malfunction. Medium priority alarms require a timely
operator response.
Low priority alarms: indicate a discomfort condition, a device
malfunction, or an improper operation. Low priority alarms
require the operator to be aware of this condition.
Prompt messages: provide additional information of the patient
or the monitor.
The monitor has preset alarm priorities for physiological alarm and
technical alarm. For more information, see
and
5.4 Alarm Mode
The monitor provides audible and visual alarm indications when an
alarm occurs. For more information, see the following table.
Alarm
Indication
High
Priority
Alarm
Medium
Priority
Alarm
Low
Priority
Alarm
Prompt
Message
Remarks
Summary of Contents for AIView 12V
Page 1: ...AIView Series Patient Monitor Operator s Manual...
Page 2: ......
Page 6: ...IV This page intentionally left blank...
Page 24: ...1 7 This page intentionally left blank...
Page 26: ...2 2...
Page 34: ...2 10 2 6 5 Bottom View 1 2 2 2 2 1 Battery compartment 2 Anti slip foot pad...
Page 35: ...2 11 This page intentionally left blank...
Page 44: ...3 9 1 2 3 4 5 8 9 7 6...
Page 55: ...3 20 This page intentionally left blank...
Page 71: ...5 11 This page intentionally left blank...
Page 78: ...6 7 This page intentionally left blank...
Page 106: ...8 5 This page intentionally left blank...
Page 112: ...9 6 This page intentionally left blank...
Page 119: ...10 7 This page intentionally left blank...
Page 129: ...11 10 2 Set PR Source select an appropriate PR source from the drop down list...
Page 165: ...15 12 Other sources of interference if any...
Page 166: ...15 13 This page intentionally left blank...
Page 194: ...17 21 This page intentionally left blank...
Page 204: ...19 5 This page intentionally left blank...
Page 210: ...20 6 This page intentionally left blank...
Page 229: ...23 10 This page intentionally left blank...
Page 248: ...25 8 This page intentionally left blank...
Page 251: ...A 3 Shutdown Delay At least 5 minutes since the first low battery alarm...
Page 283: ...D 2 This page intentionally left blank...
Page 289: ...E 6 less than 3V m...