Functions of the operating and display elements
17
Opening the water inlet
Open the lower housing cover.
Slowly open the manually operated shut-off valve to prevent the
measuring chamber overflowing. The flow regulator requires a few
seconds to function correctly.
Make sure that the water conducting parts are not leaky.
If water sprays from the hose
of the measuring chamber
,
reduce the amount of inlet water via the manually operated shut-
off valve. It should take 2 to 6 seconds to fill the measuring
chamber!
Instrument settings and data input
Please read the following information before carrying out settings and
entering data for operating the instrument.
Functions of the operating and
display elements
The Testomat 2000
®
Polymer
display shows operating statuses and
measured values. The input keys for programming (cursor block) and
the function keys are located underneath the display.
Switching Testomat 2000
®
Polymer
on/off
(1) Mains switch
Use this switch to switch the instrument on or off.
(2) Instrument fuse (inside the instrument)
This fuse protects Testomat 2000
®
Polymer and the outputs
against overloads and short circuits.
(1)
(2)