Changing the Specimen
WARNING
Risk of permanent eye damage from laser radiation
Never change specimens during the scanning operation because laser
radiation can escape uncontrolled from the specimen area.
20.1
Changing the Specimen on an Upright Microscope
To change specimens on an upright microscope, proceed as follows:
1. Finish the scanning operation.
2. Ensure that no laser radiation exists in the specimen area.
3. Replace the specimen. Insert the specimen correctly into the specimen holder.
4. If you carry out FCS measurements, you have to reset the correction ring of the objective
after every specimen change.
20.2
Changing the Specimen on an Inverted Microscope
To change specimens on an inverted microscope, proceed as follows:
1. Finish the scanning operation.
2. Ensure that no laser radiation exists in the specimen area.
3. Tilt the transmitted light arm back.
4. Replace the specimen. Insert the specimen correctly into the specimen holder.
5. Tilt the transmitted light arm back into the working position.
6. If you carry out FCS measurements, you have to reset the correction ring of the objective
after every specimen change.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......