F(L)CS Data Acquisition
Figure 166:
Settings for acquiring a reference image using photomultipliers
4. For a better comparison of the confocal image with the FCS measurement data, set the
spectral detection window for image acquisition using the internal photomultipliers
corresponding to the transmission of the SMD filter block used in the X1 port adapter.
The laser lines and intensities for image acquisition may be the same as those for the FCS
measurement, but they do not have to be.
5. Check that the external port X1 is switched to the
Mirror
position (
).
6. Start a continuous scan by clicking the
Live
image acquisition button (
)
and start an image acquisition by clicking the
Capture Image image acquisition button
).
18.1.6.2
Image Acquisition with PE APDs (AQR Type) or MPD APDs (PDM Type)
You can also acquire the image directly using the APDs. This procedure is recommended
specifically for specimens with a low signal.
General recommendations for APD image acquisition:
•
Slow scan speed (e.g. 10 to 100 Hz)
•
Data accumulation (e.g. 4 to 8 frames)
•
Best detection range for PE APDs: over 500 to approx. 800 nm, for MPD APDs:
between 430 nm and 700 nm
•
Adapt the scale of the color look-up table to the counting rate if necessary
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......