FLIM Data Acquisition
Figure 121:
Selecting external APDs for FLIM data acquisition
5. If the APDs still have not been activated in
Beam Path Settings
, click the buttons
allocated to the APDs and set them to
ON
.
6. Deactivate the internal photomultipliers. Internal photomultipliers cannot be used for
imaging simultaneously with APD detection.
7. If no image is displayed, check whether the
---
option is set at
X1 Port
(
).
8. Activate at least one detector (APD), even if no image in LAS AF is required. If no
detector is activated in LAS AF at all, it will not be possible to start a scanning operation.
16.2.1.3
FLIM Data Acquisition with External Photomultiplier
Proceed as follows to prepare the TCS SP8 for the collection of FLIM data using an external
photomultiplier:
1. Switch on the photomultiplier
electrically.
2. Select a suitable range for spectral detection by choosing the appropriate SMD filter
Dependency of the spectral detection range of the APDs
The spectral detection range of the APDs is determined by the SMD filter
cube used, not by the slider settings for the photomultipliers.
Overload protection for external APDs
APDs are deactivated when emission intensity gets too high. MPD APDs
(SPADs) have to be manually reactivated. To do so, refer to the
corresponding manuals of the APD unit or the
DSN. PE APDs activate
automatically. Reduce the incident light on the APD.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......